Calibration
4-port TRL calibration
Step 7.
to start the measurement of the calibration standard.
Reflect
Softkey
Port 1 Reflect
Port 2 Reflect
Port 3 Reflect
Port 4 Reflect
Step 8.
to start the measurement of the calibration standard.
Line/Match
Softkey
*1
1-2 Line/Match
*2
1-2 Fwd (S21)
*2
1-2 Rvs (S12)
1-3 Line/Match
1-3 Fwd (S31)
1-3 Rvs (S13)
1-4 Line/Match
1-4 Fwd (S41)
1-4 Rvs (S14)
2-3 Line/Match
2-3 Fwd (S32)
2-3 Rvs (S23)
2-4 Line/Match
146
Function
Executes reflection calibration for test port 1.
Executes reflection calibration for test port 2.
Executes reflection calibration for test port 3.
Executes reflection calibration for test port 4.
Function
Executes line/match calibration for test ports 1 and 2.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports 1 and 2.Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports 1 and 2.Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports 1 and 3.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports 1 and 3.Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports 1 and 3.Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports 1 and 4.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports 1 and 4.Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports 1 and 4.Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports 2 and 3.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports 2 and 3.Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports 2 and 3.Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports 2 and 4.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.