Measurement Examples
Executing Power Calibration
2. Setting Address of Power Meter
Follow these steps to configure the power meter's GPIB address.
Setting Description
GPIB address of the power meter: 14
3. Setting Stimulus Condition
Follow these steps to configure the stimulus conditions:
Setting Description
Presetting
Sweep type: Power
Fixed frequency: 1 GHz
Power range: -40 dBm to -8 dBm
Start value: -40 dBm
Stop value: -10 dBm
Number of points: 61
4. Executing Zero Adjustment and Calibration of Power Sensor
Execute the zero adjustment and calibration of the power sensor according to the E4418B
Power Meter User's Guide.
5. Setting Calibration Data Measurement Conditions
Follow these steps to set the calibration data measurement conditions:
Setting Description
Selecting the test port: 1
Selection of power sensor: A
Number of measurements at one
measurement point: 4
Tolerance during power calibration: 5 dB
530
Key Operation
-
-
Misc Setup
GPIB Setup
-
Address
Key Operation
-
OK
-
Sweep Type
-
-
Power
CW Freq
-
-
Power
Power Ranges
-
-
Points
Key Operation
-
Power Calibration
-
Power Calibration
-
Power Calibration
- Power Calibration - Tolerance -
-
Power Meter
-
Power Sweep
-
-
-40 to -8
-
-
Select Port
1
-
Use Sensor [ A ]
-
-
Num of Readings
Chapter 15