Optimizing Measurements
Setting Port Extensions and Loss Values

Setting loss values

In addition to port extension, you can set loss values for each port. By correcting loss due
to port extension, more accurate measurement results are obtained.
There are two types of loss value settings: loss values at two frequency points for a
specified port, and a DC loss value. You can make these settings at the same time for each
port.
NOTE
You can set loss values channel by channel. Setting loss values for one particular channel
does not affect other channels.
1. Setting loss values
Step 1. Press
values.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Press
frequency (
Step 7. If you want to set loss at two frequency points, press
(enabled), and enter a loss value (
Step 8. If you want to set loss values for other ports, repeat Step 5 to Step 7.
NOTE
When you specify two frequency points, set the lower frequency to Loss1, and the higher
one to Loss2.
Equation 12-2
Expression to calculate loss using Loss 1
Equation 12-3
Expression to calculate loss using Loss 1 and Loss 2
402
or
.
.
Port Extensions
.
Loss
to select the port for which you want to set a loss value.
Select Port
to toggle to
Loss1 [OFF]
Loss1 [ON]
).
Freq1
Loss(f)
=
Loss1
Loss(f)
=
Loss1
Loss1
log
------------- -
10
Loss2
n
=
----------------------------- -
Freq1
------------- -
log
10
Freq2
to activate the channel for which you want to set loss
(enabled), and enter a loss value (
Loss2 [OFF]
) and a frequency (
Loss2
Freq2
f
×
------------- -
Freq1
n
f
×
------------- -
Freq1
) and a
Loss1
to toggle to
Loss [ON]
).