Limit Test
Ripple Test
Figure 11-9
Ripple limit table
Step 6. Press
parameters for the frequency band:
Parameter
Type
Begin Stimulus
End Stimulus
Max Ripple
386
to add a frequency band to the ripple limit table and then specify the following
Add
Description
Selects a frequency type, either ON or OFF.
ON
OFF
Specifies the start point for the stimulus value in the ripple test.
*1*2
Specifies the end point for the stimulus value in the ripple test.
*1*2
Specifies the ripple limit value.
*1. Acceptable range for the stimulus value: -500G to +500G. If any outranging value is
specified, it will be reset to fall within the range.
*2. Even if the E5070B/E5071B's sweep range is changed after the stimulus value has been
set, the stimulus value is not susceptible.
Band used for the ripple test.
Band not used for the ripple test.
Chapter 11