that will be cleared if you select the test port and execute acquisition of the calibration
coefficient (by pressing
Step 7.
to start the measurement of the calibration standard.
Thru/Line
Softkey
Function
Executes thru/line calibration for test ports x and y.
Port x-y Thru
Step 8.
to start the measurement of the calibration standard.
Reflect
Softkey
Function
Executes reflection calibration for test port x.
Port x Reflect
Executes reflection calibration for test port y.
Port y Reflect
Step 9.
to start the measurement of the calibration standard.
Line/Match
Softkey
Function
*1
Executes line/match calibration for test ports x and y. Use this to perform
x-y Line/Match
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports x and y. Use this to perform
*2
x-y Fwd (Syx)
forward-direction measurement only.
Executes line/match calibration for test ports x and y. Use this to perform
*2
x-y Rvs (Sxy)
reverse-direction measurement only.
*1.This is equivalent to measuring
*2.Measuring
x-y Fwd (Syx)
Step 10. Press
to finish TRL 2-port calibration. At this point, the calibration coefficient is
Done
calculated and saved. The error correction function is automatically turned on.
Chapter 4
).
Done
and
x-y Fwd (Syx)
and
is equivalent to measuring
x-y Rvs (Sxy)
Calibration
2-port TRL calibration
.
x-y Rvs (Sxy)
x-y Line/Match
.
139