Measurement Throughput Summary - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Measurement Throughput Summary

Table 16-20
Typical Cycle Time for Measurement Completion
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
*1.Typical performance.
*2.Sweep mode: Fast swept. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System error
correction: OFF.
Table 16-21
Typical Cycle Time for Measurement Completion
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
*1.Typical performance.
*2.Sweep mode: Fast swept. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System error
correction: ON.
Chapter 16
Specifications and Supplemental Information
Number of Points
51
201
4
5
5
8
11
12
20
23
19
24
37
46
Number of Points
51
201
4
6
5
10
11
12
20
24
20
24
37
46
Measurement Throughput Summary
*1*2
(ms)
401
1601
7
18
13
42
13
23
25
46
24
24
48
50
*1*2
(ms)
401
1601
8
22
16
56
13
23
25
55
24
26
47
57
555

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents