Manual Changes
Manual Changes
o Assignable x-axis such as RF+LO, RF-LO, and LO-RF frequencies for each active
o Conjugation for converting vector mixer measurement parameters
o 7 mm calibration kits such as 85031B and 85050C/D
o Calibration data and calibration coefficient clear functions
Change 6
Use Following specifications for the serial prefix MY422.
Table A-3
Description
Damage Level
300 kHz to 8.5 GHz

Change 5

Use Following specifications for the serial prefix MY421 and JP1KK.
Table A-4
System Dynamic Range
Description
*1*2
System Dynamic Range
300 kHz to 3 MHz
IF bandwidth = 3 kHz
3 MHz to 1.5 GHz
1.5 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
300 kHz to 3 MHz
IF bandwidth = 10 Hz
3 MHz to 1.5 GHz
1.5 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
602
trace
Specification
Specification
95 dB
97 dB
93 dB
88 dB
81 dB
120 dB
122 dB
118 dB
113 dB
106 dB
Supplemental Information
RF +20 dBm,
±14 VDC (Source attenuator = 0
dB),
±25 VDC (Source attenuator = 5
dB or more), typical
Supplemental Information
85 dB
98 dB
100 dB
95 dB
91 dB
84 dB
110 dB
123 dB
125 dB
120 dB
116 dB
109 dB