Measurement Examples
Evaluating a Duplexer
Step 2. Enter the setup data in the limit table for trace 1 (Figure 15-16).
Figure 15-16
Completed limit table for trace 1
Step 3. Display the limit table for Trace 2 (S
Setup Description
Activating Trace 2
Step 4. Enter the setup data in the limit table for trace 2 (Figure 15-17).
Figure 15-17
Completed limit table for trace 2

11. Execute the Limit Test

The limit test is executed.
Step 1. Turn on the limit line and limit test for Trace 1.
Setup Description
Activating Trace 1
Limit Line: ON
Limit Test: ON
502
).
21
Key Operation
Key Operation
-
Limit Test
(Turn it
Limit Test
-
(Turn it
)
Limit Line
ON
)
ON