Agilent Technologies ENA Series User Manual page 413

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

5. Measuring the OPEN/SHORT standard and executing calculation
Calculate port extensions and loss values based on the calculation results using the
OPEN/SHORT standard.
Step 1. Press
port extension.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. If you use the OPEN standard, press
want to execute measurement. Execution is restricted to ports selected in "1. Selecting a
port(s)" on page 404.
Softkey
All
Port 1
Port 2
Port 3
Port 4
*1.Only with Options 313, 314, 413, and 414.
*2.Only with Options 413 and 414.
Step 6. If you use the SHORT standard, press
you want to execute measurement. Execution is restricted to ports selected in "1. Selecting
a port(s)" on page 404.
Softkey
All
Port 1
Port 2
Port 3
Port 4
*1.Only with Options 313, 314, 413, and 414.
*2.Only with Options 413 and 414.
NOTE
If a port extension value or loss value has been set, the value is updated to the calculated
result.
Chapter 12
or
.
.
Port Extensions
.
Auto Port Extension
Function
Measures the OPEN standards for all ports selected in "1. Selecting a
port(s)" on page 404.
Measures the OPEN standard for port 1.
Measures the OPEN standard for port 2.
Measures the OPEN standard for port 3.
*1
*2
Measures the OPEN standard for port 4.
Function
SHORT
Measures the
port(s)" on page 404.
SHORT
Measures the
SHORT
Measures the
SHORT
*1
Measures the
SHORT
*2
Measures the
Optimizing Measurements
Setting Port Extensions and Loss Values
to activate the channel for which you want to set auto
, and select the port(s) for which you
Measure OPEN
, and select the port(s) for which
Measure SHORT
standards for all ports selected in "1. Selecting a
standard for port 1.
standard for port 2.
standard for port 3.
standard for port 4.
407