Evaluating Transmission Characteristics of a Front End
Module
This example shows how to measure the transmission characteristics of a 6-port front end
module, as shown in Figure 15-35, by using the E5070B/E5071B and the E5091A.
Figure 15-35
Front end module

Overview of evaluation procedure

In this example, a DUT is evaluated according to the procedure shown in Table 15-9.
Table 15-9
Evaluation Procedure for 6-port Front End Module
Procedure
"1. Determining Measurement Conditions" on
page 522
"2. Setting Channel Window Allocation" on
page 522
"3. Setting the Test Ports" on page 522
"4. Setting Control Line" on page 523
"5. Setting Sweep Conditions" on page 523
"6. Setting Balance Conversion Topology" on
page 523
"7. Selecting Measurement Parameter" on
page 524
"8. Executing Calibration" on page 524
"9. Connecting DUT" on page 526
Chapter 15

Evaluating Transmission Characteristics of a Front End Module

Description
Determine the measurement conditions such as the
sweep conditions and measurement ports.
Set the allocation of the channel windows on the screen.
Determine the test port assignment for each channel.
Set the E5091A's control line.
Set the sweep range and the number of points.
Set the balance port and unbalance port assignment.
Set the measurement parameter.
Perform calibration using 4-port ECal.
Connect the DUT.