Agilent Technologies ENA Series User Manual page 117

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey
3-1 (S31)
4-1 (S41)
1-2 (S12)
3-2 (S32)
4-2 (S42)
1-3 (S13)
2-3 (S23)
4-3 (S43)
1-4 (S14)
2-4 (S24)
3-4 (S34)
NOTE
An asterisk (*) in the upper right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
calibration coefficient (pressing
Step 7. Make a THRU connection between the test ports (between the connectors to which the
DUT will be connected) selected in Step 6.
Figure 4-9
Connecting at THRU response calibration
Chapter 4
THRU Response Calibration (transmission test)
Function
Selects test port 3 (input) and test port 1 (output); Corresponds to the
determination of S31
Selects test port 4 (input) and test port 1 (output); Corresponds to the
determination of S41
Selects test port 1 (input) and test port 2 (output); Corresponds to the
determination of S12
Selects test port 3 (input) and test port 2 (output); Corresponds to the
determination of S32
Selects test port 4 (input) and test port 2 (output); Corresponds to the
determination of S42
Selects test port 1 (input) and test port 3 (output); Corresponds to the
determination of S13
Selects test port 2 (input) and test port 3 (output); Corresponds to the
determination of S23
Selects test port 4 (input) and test port 3 (output); Corresponds to the
determination of S43
Selects test port 1 (input) and test port 4 (output); Corresponds to the
determination of S14
Selects test port 2 (input) and test port 4 (output); Corresponds to the
determination of S24
Selects test port 3 (input) and test port 4 (output); Corresponds to the
determination of S34
).
Done