Softkey Functions
Analysis Menu
Key Operation
(Continued)
Limit Test
Limit Line Offset
Marker to limit offset. Centers the limit lines around the current marker
Marker ->
Amplitude
position by using the limit amplitude offset function.
Offset
Returns to the softkey display screen one level higher.
Return
Turns on/off the display of the limit test fail sign.
Fail Sign
Returns to the softkey display screen one level higher.
Return
Displays softkeys for setting up the ripple limit
Ripple Limit
Enables or disables the ripple limit test function.
Ripple Limit Test
Enables or disables the ripple limit line display.
Ripple Limit
Displays softkeys for setting up the ripple value function.
Ripple Value
Disables the display of ripple value.
OFF
Sets the absolute value as the display type of the ripple value.
ABS
Sets the margin as the display type of the ripple value.
MAR
Displays softkeys for setting up the ripple value function.
Ripple Band
Sets the value of ripple band to 1.
1
:
Sets the value of ripple band to 12.
12
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for editing the ripple limit line.
Edit Ripple Limit
Deletes the line at the cursor from the limit table.
Delete
Adds a new segment to the limit table under the line on which the cursor is
Add
located.
Displays softkeys for clearing the contents of the limit table.
Clear Limit Table
Deletes all segments in the limit table.
OK
Returns to the softkey display screen one level higher.
Cancel
Exports (saves the data in formats used by other software) the limit table to
Export to CSV
a CSV (comma-separated value) file.
File
Imports (loads the data from files in different formats) a CSV
Import from CSV
(comma-separated value) file to the limit table.
File
Returns to the softkey display screen one level higher.
Return
Turns on/off the display of the ripple limit test fail sign.
Fail Sign
Returns to the softkey display screen one level higher.
Return
Displays softkeys for setting up the bandwidth limit test function.
Bandwidth Limit
Enables or disables the bandwidth test function.
BW Test
Enables or disables the bandwidth display.
BW Display
Enables or disables the bandwidth marker display.
BW Marker
Sets the threshold of bandwidth test.
N dB Points
Sets the lower limit value.
Min Bandwidth
Sets the upper limit value.
Max Bandwidth
Turns on/off the display of the bandwide limit test fail sign.
Fail Sign
Returns to the softkey display screen one level higher.
Return
Returns to the "E5070B/E5071B Menu (Top Menu)" on page 646.
Return
660
Function
:
SCPI Command
:CALC{1-16}:LIM:OFFS:MARK
:DISP:FSIG
:CALC{1-16}:RLIM
:CALC{1-16}:RLIM:DISP:LINE
:CALC{1-16}:RLIM:DISP:VAL OFF
:CALC{1-16}:RLIM:DISP:VAL ABS
:CALC{1-16}:RLIM:DISP:VAL MAR
:CALC{1-16}:RLIM:DISP:SEL
:
:CALC{1-16}:RLIM:DISP:SEL
:CALC{1-16}:RLIM:DATA
:CALC{1-16}:RLIM:DATA
:CALC{1-16}:RLIM:DATA
:CALC{1-16}:RLIM:DATA
:MMEM:STOR:RLIM
:MMEM:LOAD:RLIM
:DISP:FSIG
:CALC{1-16}:BLIM
:CALC{1-16}:BLIM:DISP:VAL
:CALC{1-16}:BLIM:DISP:MARK
:CALC{1-16}:BLIM:DB
:CALC{1-16}:BLIM:MIN
:CALC{1-16}:BLIM:MAX
:DISP:FSIG
Appendix D