Calibration
Calibration between Ports of Different Connector Types
adapter (3 in Figure 4-38).
FWD
RVS
For example, when setting ports 2 to 4, if you want to connect port 1 of the adapter to test
port 4 of the E5070B/E5071B and port 2 of the adapter to test port 2 of the
E5070B/E5071B, select RVS as the port connection mode.
When you select Adapter File, the comment contained in the adapter file is displayed at 4
in Figure 4-38.
NOTE
If the system Z0 written in the adapter file is different from the system Z0 of the
E5070B/E5071B, "file error" is displayed at 4 in Figure 4-38. In this case, you cannot set
the E5070B/E5071B to the adapter calibration mode.
Saving and loading calibration kit settings
You can save the selection of the calibration kit and standard for each test port (the setting
in the Reflection tab) and that for each pair of test ports (the setting in the Thru tab), as well
as load them for restoring whenever needed.
Press the
Press the
(3 in Figure 4-36) to recall the setting from the file.
NOTE
If the calibration kit definition is changed after saving the file, resulting in a contradiction
between the information in the file and that in the calibration kit definition, you can no
longer recall the settings from the file.
Step 5. Press
kits.
5. Performing Calibration
Set the E5070B/E5071B to the special calibration mode in which you can use a different
calibration kit for each test port (adapter calibration mode), then perform the calibration.
Step 1. Press
adapter calibration mode.
NOTE
Do not terminate the VBA macro by force.
In the adapter calibration mode, if you terminate the VBA macro forcefully, for example,
with the
button, normal calibration can no longer be performed and the label of calibration kit 10
remains altered. To return to the normal calibration mode, restart the firmware of the
E5070B/E5071B. In this case, you cannot restore the label of calibration kit 10.
Step 2. The screen showing the selected calibration kits for each test port and each pair of test ports
158
Port 1 of the adapter (port 1 of the 2-port Touchstone file) is connected to the test port of
the smaller port number of the E5070B/E5071B.
Port 2 of the adapter (port 2 of the 2-port Touchstone file) is connected to the test port of
the smaller port number of the E5070B/E5071B.
button (5 in Figure 4-37 or 6 in Figure 4-38) to save the setting to a file.
Save
button (4 in Figure 4-37 or 5 in Figure 4-38) or the
Recall
(6 in Figure 4-37 or 7 in Figure 4-38) to finish the setting of the calibration
Close
Set Adapter Calibration Mode
key before returning to the normal calibration mode with the
(4 in Figure 4-36) to set the E5070B/E5071B to the
button
Recall Cal Kit
Exit
Chapter 4