Agilent Technologies ENA Series User Manual page 677

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Key Operation
(Continued)
Calibrate
*1
3-Port TRL Cal
Displays softkeys for executing reflection calibrations. Symbols x, y and z refer to the three test ports selected in
Reflect
the
Select Ports
Executes a OPEN/SHORT reflection calibration on test port x.
Port x
Reflect
Executes a OPEN/SHORT reflection calibration on test port y.
Port y
Reflect
Executes a OPEN/SHORT reflection calibration on test port z.
Port z
Reflect
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing LINE/MATCH calibrations. Symbols x, y, and z refer to the three test ports
Line/Match
selected in the
214).
Executes a LINE/MATCH calibration on test ports x and y.
x-y
Line/Match
Executes a forward LINE/MATCH calibration on test ports x and y.
x-y Fwd(Syx)
Executes a reverse LINE/MATCH calibration on test ports x and y.
x-y Rvs(Sxy)
Executes a LINE/MATCH calibration on test ports x and z.
x-z
Line/Match
Executes a forward LINE/MATCH calibration on test ports x and z.
x-z Fwd(Szx)
Executes a reverse LINE/MATCH calibration on test ports x and z.
x-z Rvs(Sxz)
Executes a LINE/MATCH calibration on test ports y and z.
y-z
Line/Match
Executes a forward LINE/MATCH calibration on test ports y and z.
y-z Fwd(Szy)
Executes a reverse LINE/MATCH calibration on test ports y and z.
y-z Rvs(Syz)
Returns to the softkey display screen one level higher.
Return
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
*2
Displays softkeys for executing 4-port TRL calibrations.
4-Port TRL Cal
Displays softkeys for executing THRU/LINE calibrations.
Thru/Line
Executes a THRU/LINE calibration for test port 1→2.
1-2 Thru/Line
Executes a THRU/LINE calibration for test port 1→3.
1-3 Thru/Line
Executes a THRU/LINE calibration for test port 1→4.
1-4 Thru/Line
Executes a THRU/LINE calibration for test port 2→3.
2-3 Thru/Line
Executes a THRU/LINE calibration for test port 2→4.
2-4 Thru/Line
Executes a THRU/LINE calibration for test port 3→4.
3-4 Thru/Line
Returns to the softkey display screen one level higher.
Return
Appendix D
Function
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
menu (x, y, and z are always 1, 2, and 3 respectively, for models with Option 213 or
Select Ports
Softkey Functions
Calibration Menu
SCPI Command
:SENS{1-16}:CORR:COLL:TRLR x
:SENS{1-16}:CORR:COLL:TRLR y
:SENS{1-16}:CORR:COLL:TRLR z
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:TRLL y,x
:SENS{1-16}:CORR:COLL:TRLL y,x
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:TRLL x,z
:SENS{1-16}:CORR:COLL:TRLL z,x
:SENS{1-16}:CORR:COLL:TRLL z,x
:SENS{1-16}:CORR:COLL:TRLL x,z
:SENS{1-16}:CORR:COLL:TRLL y,z
:SENS{1-16}:CORR:COLL:TRLL z,y
:SENS{1-16}:CORR:COLL:TRLL z,y
:SENS{1-16}:CORR:COLL:TRLL y,z
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
:SENS{1-16}:CORR:COLL:TRLT 1,2
:SENS{1-16}:CORR:COLL:TRLT 2,1
:SENS{1-16}:CORR:COLL:TRLT 1,3
:SENS{1-16}:CORR:COLL:TRLT 3,1
:SENS{1-16}:CORR:COLL:TRLT 1,4
:SENS{1-16}:CORR:COLL:TRLT 4,1
:SENS{1-16}:CORR:COLL:TRLT 2,3
:SENS{1-16}:CORR:COLL:TRLT 3,2
:SENS{1-16}:CORR:COLL:TRLT 2,4
:SENS{1-16}:CORR:COLL:TRLT 4,2
:SENS{1-16}:CORR:COLL:TRLT 3,4
:SENS{1-16}:CORR:COLL:TRLT 4,3