Optimizing Measurements
Improving Measurement Throughput
Turning off the updating of information displayed on the LCD screen
Turning off the updating of information displayed on the LCD screen eliminates the
processing time required to update displays within the analyzer, improving measurement
throughput. If it is not necessary to check displayed information during measurements,
turning off real-time updating is an effective means of improving throughput.
The updating of information displayed on the LCD screen can be switched using the
following procedure:
Turning off the updating of information
Step 1. Press
Step 2. Press
When the LCD screen update is turned off,
Bar" on page 44.
Turning off system error correction
The E5070B/E5071B executes "IF Range Correction" on page 731 and "Port
Characteristics Correction" on page 731 in the data processing flow shown in Figure E-2,
"Data Processing Flowchart," on page 730, by using the system calibration data set at the
factory. This system error correction process is not required if the user performs proper
calibration by using the
automatically turns on error correction.
By turning off system error correction, you can reduce the data processing time needed
during measurement and thus improve measurement throughput.
NOTE
When you turn ON/OFF system error correction, all calibration data set by user calibration
is deleted.
Procedure
Step 1. Press
Step 2. Press
Step 3. Press
Figure 12-7
Dialog box for changing system error correction
Step 4. Press
414
.
to switch the updating of displayed information on the LCD screen on/off.
Update
and the softkeys that subsequently appear, which
.
.
Service Menu
. Figure 12-7 appears.
System Correction
to turn ON/OFF system error correction.
OK
appears on "4. Instrument Status
Update Off
Chapter 12