Test Port Output (Source) - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Test Port Output (Source)

Table 16-7
Test Port Output Frequency
Description
Range
E5070B
E5071B
Resolution
Source Stability
Standard
Option 1E5
CW Accuracy
Standard
Option 1E5
Table 16-8
Test Port Output Power
Description
Level Accuracy (at 23°C±5°C)
300 kHz to 10 MHz
10 MHz to 8.5 GHz
Level Accuracy (high temperature mode: ON)
300 kHz to 8.5 GHz
Level Accuracy (swept mode)
300 kHz to 4.25 GHz
4.25 GHz to 8.5 GHz
Level Linearity (23°C±5°C)
10 MHz to 3 GHz
3 GHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
Chapter 16
Specifications and Supplemental Information
Specification
300 kHz to 3 GHz
300 kHz to 8.5 GHz
1 Hz
±5 ppm, 23°C±5°C
±1 ppm, 23°C±5°C
*1
Specification
±0.650 dB (at 0 dBm, 50 MHz
absolute, Source attenuator 0 dB)
±1.0 dB (at 0 dBm, relative to 50
MHz reference, Source attenuator
0 dB)
±0.75 dB (at −15 dBm to 10 dBm)
±0.75 dB (at −15 dBm to 9 dBm)
±0.75 dB (at −15 dBm to 7 dBm )
±0.75 dB (at −15 dBm to 5 dBm )
Test Port Output (Source)
Supplemental Information
±5 ppm (5°C to 40°C, typical)
±0.05 ppm (23°C±5°C, typical)
±0.5 ppm/year (typical)
Supplemental Information
±1.0 dB (at 0 dBm, relative to 50
MHz reference, Source attenuator
0 dB)
±0.8 dB (at 0 dBm, 50 MHz
absolute, Source attenuator 0 dB)
±1.5 dB (at 0 dBm, relative to 50
MHz reference, Source attenuator
0 dB)
±2.5 dB (at 0 dBm, relative to 50
MHz reference, Source attenuator
0 dB)
±3.5 dB (at 0 dBm, relative to 50
MHz reference, Source attenuator
0 dB)