Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Procedures for setting frequency at each port.
The frequencies set at Port-n (n: any port number) should be a multiplier (M
and offset (O
Figure 8-1):
<Port (n) Frequency> = <Basic Frequency> × M
Figure 8-1
Example of frequency setting at each port
The basic frequency is a frequency range set by the
NOTE
STIMULUS block.
Follow the steps below to set the frequency at each port.
Step 1. Press
is to be set.
Step 2. Press
Step 3. Press
Step 4. Set the frequency for Port-n by using the following softkeys:
Softkey
Multiplier
Divisor
Offset
Start
Stop
NOTE
You can use both the
314
) for the basic frequency setting shown in the following formula (refer to
n
(or
-
Frequency Offset
(n: any port number) to display "Port n Menu."
Port n
/
Multiplier
Divisor
/ D
+ O
n
n
) to activate the channel on which the frequency-offset
to display "Frequency Offset Menu."
Function
Sets a multiplier for the basic frequency
Sets a divisor for the basic frequency
Sets an offset frequency for the basic frequency
Sets a start frequency of the sweep range for Port-n
Sets a stop frequency of the sweep range for Port-n
/
key and
Offset
Start
), divisor (D
n
n
and
keys in the
/
key to set the frequency
Stop
Chapter 8
)
n