Measurement Menu (Balanced Measurement, Bal-Bal) - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey Functions

Measurement Menu (Balanced Measurement, Bal-Bal)

Measurement Menu (Balanced Measurement, Bal-Bal)
Key Operation
Displays softkeys for setting up measurement parameters (only for models with Option 412 or 414). To display
these softkeys, the balanced/unbalanced conversion topology must be set to balanced-balanced (
balanced/unbalanced conversion function must be on (
(
Fixture Simulator ON
Selects parameter S
Sdd11
signal input to (balanced) port 1 on the DUT is reflected as a
differential signal.
Selects parameter S
Sdd21
signal input to (balanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a differential signal.
Selects parameter S
Sdd12
signal input to (balanced) port 2 on the DUT is transmitted to
(balanced) port 1 on the DUT as a differential signal.
Selects parameter S
Sdd22
signal input to (balanced) port 2 on the DUT is reflected as a
differential signal.
Selects parameter S
Scd11
input to (balanced) port 1 on the DUT is reflected as a common
mode signal.
Selects parameter S
Scd21
signal input to (balanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a common mode signal.
Selects parameter S
Scd12
signal input to (balanced) port 2 on the DUT is transmitted to
(balanced) port 1 on the DUT as a common mode signal.
Selects parameter S
Scd22
signal input to (balanced) port 2 on the DUT is reflected as a
common mode signal.
Selects parameter S
Sdc11
signal input to (balanced) port 1 on the DUT is reflected as a
differential signal.
Selects parameter S
Sdc21
signal input to (balanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a differential signal.
Selects parameter S
Sdc12
signal input to (balanced) port 2 on the DUT is transmitted to
(balanced) port 1 on the DUT as a differential signal.
Selects parameter S
Sdc22
signal input to (balanced) port 2 on the DUT is reflected as a
differential signal.
Selects parameter S
Scc11
signal input to (balanced) port 1 on the DUT is reflected as a
common mode signal.
Selects parameter S
Scc21
signal input to (balanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a common mode signal.
Selects parameter S
Scc12
signal input to (balanced) port 2 on the DUT is transmitted to
(balanced) port 1 on the DUT as a common mode signal.
704
Function
) in the "Analysis Menu" on page 647.
. S
defines the way a differential
dd11
dd11
. S
defines the way a differential
dd21
dd21
. S
defines the way a differential
dd12
dd12
. S
defines the way a differential
dd22
dd22
. S
defines the way a differential signal
cd11
cd11
. S
defines the way a differential
cd21
cd21
. S
defines the way a differential
cd12
cd12
. S
defines the way a differential
cd22
cd22
. S
defines the way a common mode
dc11
dc11
. S
defines the way a common mode
dc21
dc21
. S
defines the way a common mode
dc12
dc12
. S
defines the way a common mode
dc22
dc22
. S
defines the way a common mode
cc11
cc11
. S
defines the way a common mode
cc21
cc21
. S
defines the way a common mode
cc12
cc12
SCPI Command
), and the fixture simulator function must be on
BalUn ON
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDD11
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDD21
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDD12
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDD22
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCD11
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCD21
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCD12
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCD22
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDC11
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDC21
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDC12
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SDC22
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCC11
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCC21
:CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL
SCC12
), the
Bal-Bal
Appendix D

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents