Corrected System Performance

The specifications in this section apply for measurements made with the Agilent
E5070B/E5071B Network Analyzer under the following conditions:
No averaging applied to data
Environmental temperature of 23°C ±5°C, with less than 1°C deviation from the
calibration temperature
Response and isolation calibration not omitted
Table 16-1
System Dynamic Range
Description
*1*2
System Dynamic Range
300 kHz to 3 MHz
IF bandwidth = 3 kHz
3 MHz to 1.5 GHz
1.5 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
300 kHz to 3 MHz
IF bandwidth = 10 Hz
3 MHz to 1.5 GHz
1.5 GHz to 3GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
*1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source
maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals
into account.
*2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.25GHz due to spurious receiver
residuals.
Chapter 16
Specifications and Supplemental Information
Corrected System Performance
Specification
95 dB
97 dB
96 dB
92 dB
87 dB
80 dB
120 dB
122 dB
121 dB
117 dB
112 dB
105 dB
Supplemental Information
85 dB
98 dB
100 dB
99 dB
94 dB
90 dB
83 dB
110 dB
123 dB
125 dB
124 dB
119 dB
115 dB
108 dB
535