Step 10. Press
screen.
Step 11. Press
Step 12. Repeat Step 6 to Step 11 to redefine all standards for which changes are necessary.
Step 13. Press
Step 14. Press
Step 15. Select the class to be redefined according to the list below.
Softkey
Sub Class
Open
Short
Load
Thru
TRL Reflect
TRL Line/Match
Step 16. For TRL Thru, select ports.
Softkey
Port 1-2
Port 1-3
Port 1-4
Port 2-3
Port 2-4
Port 3-4
NOTE
Select
Step 17. Select the test port according to the list below.
Softkey
Port 1
Port 2
Port 3
Port 4
Chapter 4
and input a new label for the standard using the keypad displayed on the
Label
.
Return
.
Return
.
Specify CLSs
Function
Selects a subclass you want to use.
Selects the OPEN class
Selects the SHORT class
Selects the LOAD class
Selects the THRU class
Selects TRL Reflect.
Selcts TRL Line/Match.
Function
Selects port 1 and port 2.
Selects port 1 and port 3.
Selects port 1 and port 4.
Selects port 2 and port 3.
Selects port 2 and port 4.
Selects port 3 and port 4.
to use the same standards for all test ports.
Set All
Function
Selects port 1
Selects port 2
Selects port 3
Selects port 4
Changing the Calibration Kit Definition
Calibration
187