Agilent Technologies ENA Series User Manual page 170

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration
Adapter Characterization
6. Measuring Data
Measure data when each standard is connected.
Step 1. Connect the adapter to test port 1 of the E5070B/E5071B.
Step 2. According to the setting of Cal Kit 1, connect each standard to the adapter and then press
the corresponding button (4 in Figure 4-41). When the data measurement is complete, the
button turns yellow.
When you calculate the S-parameters using measurement data in one direction only, the
data measurement is complete here.
Step 3. Connect the adapter to test port 2 of the E5070B/E5071B in the reverse direction.
Step 4. According to the setting of Cal Kit 2, connect each standard to the adapter and then press
the corresponding button (5 in Figure 4-41). When the data measurement is complete, the
button turns yellow.
7. Saving to a File
Calculate the S-parameters and save them to a file.
Step 1. Press the
button (6 in Figure 4-41).
Save
Step 2. The Save As dialog box appears. Enter a file name and press
If you save the file under the D:\Agilent\Data\AdapterCharacterization folder, you can
select it as the standard Adapter file while setting the adapter calibration mode in the
Specify Cal Kit dialog box.
8. Closing VBA Macro
Press
(7 in Figure 4-41) to close the macro.
Close
164
.
Save
Chapter 4