Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
Table F-5
Functions of the 8753ES vs. the E5070B/E5071B
Function
Calibration
Calibration kit
Calibration
type
Omission of the isolation measurement
Power meter calibration
Receiver calibration
Adapter removal
Velocity factor
Characteristic impedance of the measurement
system
Port extension
Selection between chop sweep and alternate
sweep
Error correction
Interpolating the calibration coefficient
748
Not more than two ports
Not less than three ports
8753ES
Available calibration kits:
7 mm: 85031B
3.5 mm: 85033C/D/E
N type (50 Ω): 85032B/E/F
N type (75 Ω): 85036B/E
2.4 mm: 85056A/D
TRL 3.5 mm: 85052C
User defined calibration kit
Response calibration
Response and isolation calibration
Enhanced response calibration
S11 1-port calibration
S22 1-port calibration
Full 2-port calibration
TRL*/LRM* Calibration
Receiver calibration
Power meter calibration
Not available
Possible
(Can be omitted by designating it
using the softkeys on the front panel
or sending a GPIB command from
the front panel)
Available
Available
Available
Definable
Definable
Definable for ports 1 and 2 and
inputs A and B
Can be changed over.
On/Off setting is definable.
On/Off setting is definable.
E5070B/E5071B
7 mm: 85031B, 85050C/D
3.5 mm: 85033D/E, 85052D
N type (50 Ω): 85032B/F
N type (75 Ω): 85036B/E
7-16 (50 Ω): 85038A/F/M
Up to 20 user defined calibration kits
can be registered.
Response calibration
(OPEN,SHORT, THRU)
Enhanced response calibration
Full 1-port calibration
Full 2-port calibration
TRL/LRM Calibration
TRL Calibration
Unknown Thru
The response calibration includes
isolation measurement as an option.
Full 3-port calibration (Opts. 313,
314, 413, and 414 only)
Full 4-port calibration (Opts. 413 and
414 only)
TRL/LRM Calibration
TRL Calibration
Unknown Thru
Possible
(Isolation measurement is performed
as an option accompanying a
response calibration and 2-, 3-, and
4-port calibration.)
Definable with fixture simulator
Definable for each port per channel
Chop sweep for traces in a channel
and alternate sweep between
channels
Always On.
Appendix F