Figure 6-10
Target search (when transition type is set to "both positive and negative")
Executing a Target Search
Step 1. Following Step 1 to Step 4 in "Reading values on the trace" on page 253, activate the
marker you are using for the target search.
Step 2. Press
Step 3. Press
Step 4. Press
This causes the target search to be executed based on the definitions of the newly set target
value and the currently set transition type.
Step 5. Press
Step 6. Select a transition type.
Softkey
Positive
Negative
Both
This sets the target search to be executed based on the definitions of the currently set target
value and the newly set transition type.
Step 7. Press the corresponding softkey to move the marker to the target.
Softkey
Search Target
Search Left
Search Right
NOTE
When the data format is in Smith chart or polar format, execute the search for the main
response value of the two marker response values.
Chapter 6
Searching for Positions that Match Specified Criteria
.
.
Target
and enter the target value in the entry area that appears.
Target Value
.
Target Transition
Function
Selects Positive as transition type
Selects Negative as transition type
Selects Both Positive and Negative as transition type
Function
Executes target search
Executes left search
Executes right search
Data Analysis