Agilent Technologies ENA Series User Manual page 135

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Figure 4-22
Connecting ECal module (full 2-port calibration)
Step 3. Press
calibration.
Step 4. Press
Step 5. Press
Step 6. To enable isolation calibration, press
Step 7. Press
this key performs a 2-port ECal.
Step 8. When using a 3-port or 4-port E5070B/E5071B (Options 313, 314, 413, or 414), press one
of the softkeys below to start a full 2-port calibration.
Softkey
Port 1-2
Port 1-3
Port 1-4
Port 2-3
Port 2-4
Port 3-4
*1.Only with Options 413 and 414.
Chapter 4
or
.
.
ECal
. When using a 2-port E5070B/E5071B (Options 213 or 214), pressing
2 Port ECal
Function
Performs a full 2-port calibration between test ports 1 and 2
Performs a full 2-port calibration between test ports 1 and 3
*1
Performs a full 2-port calibration between test ports 1 and 4
Performs a full 2-port calibration between test ports 2 and 3
Performs a full 2-port calibration between test ports 2 and 4
*1
Performs a full 2-port calibration between test ports 3 and 4
*1
ECal (electronic calibration)
to select the channel for which you want to perform the
and confirm that the display has turned
Isolation
Calibration
.
ON
129