Agilent Technologies ENA Series User Manual page 674

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
*1
3-Port Cal
Transmission
Executes a THRU calibration on test ports y and z.
Port y-z Thru
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing isolation calibrations. Symbols x, y, and z refer to the three test ports selected in
Isolation
the
(Optional)
Select Ports
Executes an isolation test on test ports x and y.
Port x-y Isol
Executes an isolation test on test ports x and z.
Port x-z Isol
Executes an isolation test on test ports y and z.
Port y-z Isol
Returns to the softkey display screen one level higher.
Return
Terminates a calibration process and calculates the calibration
Done
coefficients.
Terminates partial overwrite and recalculates the calibration
Overwrite
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing full 4-port calibrations.
*2
4-Port Cal
Displays softkeys for executing reflection calibrations.
Reflection
Executes an OPEN calibration on test port 1.
Port 1 Open
Executes a SHORT calibration on test port 1.
Port 1 Short
Executes a LOAD calibration on test port 1.
Port 1 Load
Executes an OPEN calibration on test port 2.
Port 2 Open
Executes a SHORT calibration on test port 2.
Port 2 Short
Executes a LOAD calibration on test port 2.
Port 2 Load
Executes an OPEN calibration on test port 3.
Port 3 Open
Executes a SHORT calibration on test port 3.
Port 3 Short
Executes a LOAD calibration on test port 3.
Port 3 Load
Executes an OPEN calibration on test port 4.
Port 4 Open
Executes a SHORT calibration on test port 4.
Port 4 Short
Executes a LOAD calibration on test port 4.
Port 4 Load
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing transmission calibrations.
Transmission
Executes a THRU calibration between test ports 1 and 2.
Port 1-2 Thru
Executes a THRU calibration between test ports 1 and 3.
Port 1-3 Thru
Executes a THRU calibration between test ports 1 and 4.
Port 1-4 Thru
668
Function
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
SCPI Command
:SENS{1-16}:CORR:COLL:THRU y,z
:SENS{1-16}:CORR:COLL:THRU z,y
:SENS{1-16}:CORR:COLL:ISOL x,y
:SENS{1-16}:CORR:COLL:ISOL y,x
:SENS{1-16}:CORR:COLL:ISOL x,z
:SENS{1-16}:CORR:COLL:ISOL z,x
:SENS{1-16}:CORR:COLL:ISOL y,z
:SENS{1-16}:CORR:COLL:ISOL z,y
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:PART:SAVE
:SENS{1-16}:CORR:COLL:CLE
:SENS{1-16}:CORR:COLL:OPEN 1
:SENS{1-16}:CORR:COLL:SHOR 1
:SENS{1-16}:CORR:COLL:LOAD 1
:SENS{1-16}:CORR:COLL:OPEN 2
:SENS{1-16}:CORR:COLL:SHOR 2
:SENS{1-16}:CORR:COLL:LOAD 2
:SENS{1-16}:CORR:COLL:OPEN 3
:SENS{1-16}:CORR:COLL:SHOR 3
:SENS{1-16}:CORR:COLL:LOAD 3
:SENS{1-16}:CORR:COLL:OPEN 4
:SENS{1-16}:CORR:COLL:SHOR 4
:SENS{1-16}:CORR:COLL:LOAD 4
:SENS{1-16}:CORR:COLL:THRU 1,2
:SENS{1-16}:CORR:COLL:THRU 2,1
:SENS{1-16}:CORR:COLL:THRU 1,3
:SENS{1-16}:CORR:COLL:THRU 3,1
:SENS{1-16}:CORR:COLL:THRU 1,4
:SENS{1-16}:CORR:COLL:THRU 4,1