Procedures for setting external signal source GPIB address
Follow the steps below to set GPIB address of the external signal source.
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Type the GPIB address of the external signal source you want to use.
Procedures for selecting external signal source
The E5070B/E5071B can use the external signal sources shown in Table 8-2.
Table 8-2
Available External Signal Sources
Type
1
2
3
NOTE
For type 1, the user-defined commands can be used to control the external signal source.
Four commands can be defined: preset, turning on RF output, setting frequency, and setting
power. Definitions are required for frequency and power level settings.
The factory-state definitions are shown in Table 8-3.
Table 8-3
Factory-state Command Definitions
Function
Preset
Turning on RF output
Frequency setting
Power level setting
Follow the steps below to select an external signal source.
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Chapter 8
.
.
Misc Setup
.
GPIB Setup
.
Signal Generator Address
.
Address
Model
User-defined Commands
8643A, 8644B, 8664A, 8665A/B
8648A/B/C/D, ESG Series, PSG Series
Command
""
"R3"
"FR %f% HZ"
"AP %p% DM"
.
.
Misc Setup
.
GPIB Setup
.
Signal Generator Address
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Description
Not defined
Frequency (Hz) is set to %f%
Power level (dBm) is set to %p%
317