Figure 11-10
Example of ripple limit configuration (for limit table in Figure 11-9)
NOTE
The individual frequency bands for the ripple test can overlap each other; in this case, the
ripple limit test is performed for each frequency band.
NOTE
Even if the E5070B/E5071B's span value is set to zero, you must enter a parameter for both
Begin Stimulus and End Stimulus.
NOTE
If the data format is Smith chart or polar, the limit test is performed for the main response
value among the two marker response values.
Saving/recalling ripple limit table
The ripple limit table can be saved in a file and recalled later for use on the screen. The file
is saved in the csv format (with the extension *.csv), and values are saved as a character
string with the unit. The csv formatted file can also be reused in spreadsheet software made
for PCs.
The ripple limit table is saved in the following format:
Example 11-2
Sample Ripple Limit Table Saved in CSV Format
"# Channel 1"
"# Trace 1"
Type,
ON,
ON,
ON,
Chapter 11
Begin Stimulus,
End Stimulus,
933.0000000 MHz,
964.0000000 MHz,
938.0000000 MHz,
953.0000000 MHz,
953.0000000 MHz,
960.0000000 MHz,
Limit Test
Ripple Test
MaxRipple
1.5 dB
500 mdB
300 mdB
387