Agilent Technologies ENA Series User Manual page 237

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Set the stimulus conditions for the channel you want to calibrate. For the necessary steps,
refer to "Setting Stimulus Conditions" on page 70.
You must also set the external signal source in advance. For the necessary steps, refer to "2.
Setting External Signal Source" on page 316.
NOTE
If you characterize a calibration mixer with an IF filter, we recommend that you perform
full 4-port calibration in advance, since it simplifies the evaluation procedures. For detailed
information on full 4-port calibration, see "Full 4-Port Calibration" on page 124.
Step 1. Press
Step 2. Press
Step 3. Specify "D:\Agilent\MixerCharacterization.vba" for the file name in the Open dialog box
and then press the
Step 4. Press
Step 5. Select
case, the data are saved to a temporary file (MIXER_1.s2p).
NOTE
For detailed information on characterizing the calibration mixer, see "Characterizing
procedure for calibration mixer (with IF filter)" on page 225.
Step 6. Select
data are also saved to a temporary file (MIXER_2.s2p).
NOTE
For detailed information on characterizing the calibration mixer, see "Characterizing
procedure for calibration mixer (with IF filter)" on page 225.
Step 7. Press
Figure 4-72
Vector Mixer Characterization Macro
Step 8. As the Vector Mixer Characterization Macro is running, the data files of the pre-measured
calibration mixer with IF filter (MIXER_1.s2p, MIXER_2.s2p) are read automatically into
Chapter 4
.
.
Load & Project
button.
Open
to run the macro (see Figure 4-72).
(1 in Figure 4-72) to characterize the calibration mixer 1 with IF filter. In this
Port 1
Figure 4-72) to characterize calibration mixer 2 with IF filter. Here, the
Port 2 (1 in
(2 in Figure 4-72).
(Optional) Balanced Mixer
Calibration
Vector-Mixer Calibration
231