Setting and Using the Control and Management Functions
System Recovery

Procedure to execute the factory function (1)

This procedure is valid for the E5070B : JP1KKxxxxx, MY42300632 and below, E5071B :
JP1KKxxxxx, MY42301396 and below. If your unit doesn't follow the procedure written
below, please refer to "Procedure to execute the factory recovery function (2)" on page 461
for the detail.
This section describes how to return the contents of the C drive to the factory state.
NOTE
You need the keyboard for this operation.
Step 1. Shut down the E5070B/E5071B.
Step 2. Connect the keyboard to the E5070B/E5071B.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 4. Press the standby switch of the E5070B/E5071B to turn it on.
Step 5. When the screen as shown in the figure below appears, press and hold
keyboard until this screen disappears.
NOTE
After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.