Remote Control Using HTTP
You can access the web page installed in the E5070B/E5071B by using the hypertext
transfer protocol (http) and the E5070B/E5071B's IP address from the external PC's web
browser. Through the built-in web page, you can control the E5070B/E5071B remotely and
display the measurement screen on external PCs.
The following browsers are recommended:
•
Java Script enabled browser
•
Internet Explorer 5
•
Netscape 4.7
The following information will help you to remotely control the E5070B/E5071B via the
built-in web page.
Required Modification of Settings
Network
Configure the E5070B/E5071B's network correctly. For detailed information on
configuration and notes, see "Configuring the Network" on page 434.
Enabling Web Server
Enable the web server for the E5070B/E5071B so that it may allow access from an external
PC. Follow these steps:
Step 1. Press
Step 2. Press
Step 3. Press
NOTE
This function is available only when the Internet Information Services (IIS) is installed in
the E5070B/E5071B. The web server softkey is not displayed if the IIS is not installed.
NOTE
The external PC must have the Java™ Runtime Environment installed.
How to Start VNC Server Configuration
To restrict external access, set a password for the VNC server configuration.
The following is a description of how to start the VNC server configuration. Visit the web
site at http://www.realvnc.com for information on the password setting procedure and
VNC server.
Step 1. Press
Step 2. Press
Step 3. Press
Chapter 13
Setting and Using the Control and Management Functions
.
Misc Setup - Network Setup
to turn it on. The default is OFF.
Web Server
.
Misc Setup - Network Setup
to start the VNC Server Properties.
VNC Server Configuration
Remote Control Using HTTP
.
.
439