Troubleshooting
Warning Message
This message appears when you set the sweep type to power sweep, power slope function
to enabled (ON, correction coefficient is other than zero), power calibration to ON,
frequency offset to enabled, spurious avoidance of frequency offset to enabled, or external
signal source control to enabled, while the sweep mode is set to swept mode or high speed
swept mode.
When this message is displayed, the sweep mode is automatically set to step mode (during
swept mode) or high speed step mode (during high speed swept mode).
Target value not found
This warning message is displayed when, with the tracking turned on, the target specified
by the marker search has not been found by the time the sweep is finished (with the
tracking executed).
This warning message is displayed also when, with the bandwidth marker displayed, the
setting for the bandwidth marker is changed at the end of the sweep, or when, with the
active marker changed or moved, the bandwidth is not found.
Transform, Gate not allowed
This message appears when you have set the measurement points to two, sweep type to
other than linear sweep, or frequency offset to enabled, while the gating/conversion
function of the time domain feature is turned on.
If this message is displayed, the gating/conversion function is automatically turned off.
User Preset File not found
This warning message is displayed if no user preset file (D:\UserPreset.sta) exists when
executing the SYST:UPR command.
If this message is displayed, a normal preset (SYST:PRES) will be performed.
For details on setting the user preset function, see"Setting the user preset function" on
page 449.