Optimizing Measurements
Setting Port Extensions and Loss Values
3. Specifying a loss value as a calculation target
Specify whether you want to include a loss value in the calculation result.
Step 1. Press
port extension.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Softkey
ON
OFF
NOTE
If the setting is not made before starting the measurement of the OPEN/SHORT standard, it
does not affect the calculation result.
4. Specifying a DC loss value as a calculation target
Specify whether you want to include a DC loss value in the calculation result.
Step 1. Press
port extension.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Softkey
ON
OFF
NOTE
If the setting is not made before starting the measurement of the OPEN/SHORT standard, it
does not affect the calculation result.
406
or
.
.
Port Extensions
.
Auto Port Extension
to turn it on.
Include Loss
Function
Calculates a loss value.
Does not calculate a loss value.
or
.
.
Port Extensions
.
Auto Port Extension
to turn it on.
Adjust Mismatch
Function
Calculates a DC loss value.
Does not calculate a DC loss.
to activate the channel for which you want to set auto
to activate the channel for which you want to set auto
Chapter 12