Figure 4-63
Connection of power sensor
Step 8. Press
NOTE
If the power meter GPIB address is not set correctly or if the power sensor is not connected
to the specified channel, an error occurs and calibration data are not measured.
You can abort the measurement by pressing
When the measurement is complete, the power level error correction is automatically
turned on.
NOTE
If appropriate calibration data cannot be obtained, an error may occur for each sweep after
the measurement of calibration data is complete. In this event, turn off the power level
error correction, check the connection and setting, and then measure the calibration data
again.
Chapter 4
to start the measurement of calibration data.
Take Cal Sweep
Power Calibration
during measurement.
Abort
Calibration
219