Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep

3. Select the Segment Sweep as the Sweep Type

The segment sweep is selected as the sweep type.
Setup Description
Sweep type: Segment sweep

4. Execute the Calibration

In this step, a 2-port ECal is executed on the two ports to be used.
Step 1. Connect the ECal module across test ports 1 and 2.
Figure 14-3
Connecting the ECal Module
Step 2. Execute the 2-port ECal.
Setup Description
Executing a 2-port ECal between test ports 1
and 2

5. Connect the DUT

The DUT is connected across test ports 1 and 2.