Agilent Technologies ENA Series User Manual page 242

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration
Scalar-Mixer Calibration
Softkey
1-4 (rev)
1, 4 (both)
3-2 (fwd)
2-3 (rev)
2, 3 (both)
4-2 (fwd)
2-4 (rev)
2, 4 (both)
4-3 (fwd)
3-4 (rev)
3, 4 (both)
NOTE
For the scalar-mixer calibration, only one direction with 2-ports calibration is available.
"One direction" means the forward or reverse direction. Two directions may be used
among the same ports, but this would not be full 2-port scalar-mixer calibration but simply
the simultaneous operation of single-direction scalar-mixer calibrations.
Step 7. The following steps are provided for the case where you have selected both
Figure 4-76
Connection of standards in scalar-mixer calibration
Step 8. Press
236
Function
Selects reverse direction for test port 1 and 4
Selects both directions for test port 1 and 4
Selects forward direction for test port 2 and 3
Selects reverse direction for test port 2 and 3
Selects both directions for test port 2 and 3
Selects forward direction for test port 2 and 4
Selects reverse direction for test port 2 and 4
Selects both directions for test port 2 and 4
Selects forward direction for test port 3 and 4
Selects reverse direction for test port 3 and 4
Selects both directions for test port 3 and 4
.
Reflection
and
.
x
y
Chapter 4