Frequency-Offset Measurement (Option 008)
Measurement of Harmonic Distortion
Table 8-5
Segment Table Settings
Center
3 GHz
3.5 GHz
Step 8. Set
Step 9. Press
Step 10. Press
Figure 8-15
Example of Harmonic Distortion Measurement
If you use the network analyzer as a tuning receiver like a spectrum analyzer, the setting
NOTE
must be the same at both the frequency point at which the harmonic is generated and the
measurement point. Spectrum analyzers can detect all of the signals residing in the range of
sweep frequencies, but network analyzers can only detect the signals residing in the
measurement frequency point.
For additional information about the absolute measurements, see Agilent application note
NOTE
1463-6 "Accurate Frequency-Translating Device Measurements by using the
Frequency-Offset Mode, 5989-1420EN".
330
to
Segment Display
Freq Base
to select
Sweep Type
Segment
to use Auto Scale All for trace scale optimization (Figure 8-15).
Span
1 MHz
0 Hz
.
.
Point
5
1
Chapter 8