Step 4. The dialog box shown in Figure 4-50 is displayed to confirm execution. Press
Figure 4-50) to start storing the user characteristics.
CAUTION
Do not disconnect the USB cable or terminate the VBA macro by force while the VBA
macro is storing data to the ECal's built-in flash memory. Doing so may damage the
ECal module.
Figure 4-50
Execution Confirmation screen
The dialog box shown in Figure 4-51 appears while the VBA macro is storing data to
memory. Storing the user characteristics takes a few minutes depending on the amount of
data.
Figure 4-51
Screen while macro is storing data
Step 5. Another dialog box is displayed to notify completion of data storage. Click
8. Closing the VBA macro
Press
Chapter 4
(3 in Figure 4-47) to close the macro.
Close
Calibration
User-characterized ECal
(1 in
OK
.
OK
175