Advertisement

Measurement Examples
Evaluating a Duplexer
Figure 15-9
EcalAssistant (port/channel selection) dialog box
Step 5. Following the instructions in the dialog box, select the type of ECal, test ports, and the
channel.
Setup Description
Select Ports
Select Channel
Channel 1
Step 6. Press the
Figure 15-10
EcalAssistant (connection) dialog box
Step 7. Connect the ECal module to the test ports according to the instructions in the dialog box
and then press
Setup Description
Executing calibration between ports 1 and 2
Executing calibration between ports 1 and 3
Executing calibration between ports 2 and 3
496
ECal type: Full 3-port calibration
Test ports to be used for ECal: 1, 2, and 3
Channel on which ECal is to be executed:
button. The EcalAssistant (connection) dialog box (Figure 15-10) appears.
Next
.
Measure
Key Operation
3 Port
1, 2, 3
:
Channel
1
Key Operation
(After making the connections shown in Figure
15-11 (a)), press
-
Measure
Next
(After making the connections shown in Figure
15-11 (b)), press
-
Measure
Next
(After making the connections shown in Figure
15-11 (c)), press
-
Measure
Next
Chapter 15