Performing Measurement

Trigger state and switching the setting of the E5091A

The following table shows how the setting in the E5091A is switched from when the
trigger state is the stop state. For more information on the trigger state, refer to
E5070B/E5071BProgrammers Guide.
Trigger state
Switching the setting of E5091A
Stop
The setting is not switched.
Trigger wait
The setting of the internal switch and the output of the control
line are switched according to the setting of the channel swept
first.
The connections of the test ports and the output of the control
line are switched according to the setting of the channel swept
first.
Measurement
Measurement is performed following the procedure below.
Execute a sweep for the first channel.
Set the connections of the test ports and the output of the control
line according to the settings of the channel swept second.
Execute a sweep for the second channel.
Set the connections of the test ports and the output of the control
line according to the setting of the channel swept last.
Execute a sweep for the last channel.
Stop or trigger wait
The setting is not switched for the stop state; it is switched for
the trigger wait state.

Operation

Perform operations by referring to Chapter 5, "Making Measurements," on page 241.
Chapter 14
Performing Measurement
Controlling E5091A