Performing Measurement
Trigger state and switching the setting of the E5091A
The following table shows how the setting in the E5091A is switched from when the
trigger state is the stop state. For more information on the trigger state, refer to
E5070A/E5071AProgrammers Guide.
Trigger state
Switching the setting the E5091A
Stop
The setting is not switched.
Trigger wait
The setting of the internal switch and the output of the control
line are switched according to the setting of the channel swept
first.
The connection of the test ports and the output of the control line
are switched according to the setting of the channel swept first.
Measurement
Measurement is performed following the procedure below.
Execute a sweep for the first channel.
Set the connection of the test ports and the output of control line
according to the setting of the channel swept second.
Execute a sweep for the second channel.
Set the connection of the test ports and the output of control line
according to the setting of the channel swept last.
Execute a sweep for the last channel.
Stop or trigger wait
The setting is not switched for the stop state; it is switched for
the trigger wait state.
Operation
Perform operation, referring to Chapter 5, "Making Measurements," on page 121.
13
Performing Measurement
¯
¯
¯
¯
¯
Controlling E5091A