Calibration

7
Calibration
56
The T10FS torque flange delivers an electrical calibration
signal that can be switched at the amplifier end for
measurement chains with HBM components. The
measurement flange generates a calibration signal of
about 50% of the nominal (rated) torque. The precise
value is specified on the type plate. Adjust the amplifier
output signal to the calibration signal supplied by the
connected torque flange to adapt the amplifier to the
measurement flange. To obtain stable conditions, the
calibration signal should only be activated once the
transducer has been warming up for 15 minutes.
Notice
The measurement flange should not be under load when
the calibration signal is being measured, since the
calibration signal is mixed additively.
Important
To maintain measurement accuracy, the calibration
signal should be connected for no more than 5 minutes.
A similar period is then needed as a cooling phase before
triggering the calibration signal again.
7.1

Calibration Option 2, code KF1

Increasing the excitation voltage from 54V
(pins 6 and 7, connector 1), triggers the calibration
signal.
A0785-150 HBM: public