Thermo Scientific Partisol 2000i Instruction Manual

Thermo Scientific Partisol 2000i Instruction Manual

Air sampler
Table of Contents

Advertisement

Quick Links

Partisol 2000i Air Sampler/
Partisol 2000i-D Dichotomous
Air Sampler
Instruction Manual
Part Number 110735-00
14Apr2015

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Partisol 2000i and is the answer not in the manual?

Questions and answers

Summary of Contents for Thermo Scientific Partisol 2000i

  • Page 1 Partisol 2000i Air Sampler/ Partisol 2000i-D Dichotomous Air Sampler Instruction Manual Part Number 110735-00 14Apr2015...
  • Page 2 © 2011 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin, MA 02038 1-508-520-0430 www.thermo.com/aqi...
  • Page 3 WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them.
  • Page 5 Chapter 1 “Introduction” introduces the user to the advanced features ● of the Partisol 2000i and Partisol 2000i-D and describes the flow and sampling configurations of the system. It also discusses the flow control scheme used in the unit.
  • Page 6 “AK Protocol” provides a description of the AK protocol ● commands that can be used to remotely control a Partisol 2000i and 2000i-D using a host device such as a PC or a datalogger. Safety Review the following safety information carefully before using the Sampler.
  • Page 7 A properly grounded antistatic wrist strap must be worn while handling any internal component. If an antistatic wrist strap is not available, be sure to touch the instrument chassis before Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 8 ▲ Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 9 AMTIC web site: http://www.epa.gov/ttn/amtic/pmqa.html, the QC coordinator at any EPA Regional Office, or from the Monitoring and Quality Assurance Group (MD- 14), U.S. EPA, Research Triangle Park, NC 27711. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 10 Chinook Engineering, LLC.   VSCC and Very Sharp Cut Cyclone are trademarks of BGI, Inc. Some content of this manual is courtesy of BGI, Inc. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 11 Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 12 About This Manual Where to Get Help viii Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 13: Table Of Contents

    Partisol 2000i Inlet Conversion ..............4-1     Partisol 2000i PM VSCC Components ..........4-1   Partisol 2000i PM WINS Components ..........4-2   Partisol 2000i PM Components ............4-3   Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 14 Contents Converting a Partisol 2000i from a PM Sampler to a PM Sampler . 4-4   Converting a Partisol 2000i from a PM Sampler to a PM Sampler 4-7   Chapter 5 Operational Modes ................... 5-1     Partisol Operational Modes ..............5-1  ...
  • Page 15 Contents Flow Audit/Verification – Partisol 2000i-D ........7-8   Internal Leak Check ............... 7-10   Programming a Sample Run ............. 7-11   Post-Sampling Verification and Data Retrieval ........7-14   Viewing and Downloading Stored Data ......... 7-15   USB Data Download ..............7-18  ...
  • Page 16   Reset to Default Content ............9-32   Configure Datalogging ..............9-33   Logging Period Min ..............9-33   Memory Allocation Percent ............9-33   Data Treatment ................9-34   Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 17 Unlock Instrument ..............9-52   Change Password ................ 9-52   Remove Password ................ 9-52   Restore Defaults ................9-53   Audit and Calibration Menu ............. 9-53   Audit Mode ................... 9-53   Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual xiii...
  • Page 18 Number of Items ................. 9-73   Item Number ................9-73   Other Measurements ..............9-73   Other Measurements 2 ..............9-74   Analog Inputs ................9-75   Non-Measurements ..............9-76   Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 19 Contact Information ..............9-94   Chapter 10 Preventive Maintenance ................10-1     Safety Precautions ................10-1   Routine Maintenance Items .............. 10-1   Partisol 2000i Maintenance Items ..........10-2   Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 20 WINS Impactor Removal .............. 10-9   WINS Impactor Cleaning ............10-10   WINS Impactor Reinstallation ............. 10-12   Virtual Impactor Maintenance Partisol 2000i-D ......10-12   Virtual Impactor Cleaning Partisol 2000i-D ........ 10-13   V-Seals Cleaning and Replacement ..........10-15  ...
  • Page 21   TCP Communication Parameters ............C-2   Application Data Unit Definition ............C-2   Slave Address ................... C-2   MBAP Header ................C-2   Function Code ................C-3   Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual xvii...
  • Page 22 Multiple Instruments of the Same Model ........D-27   Multiple Instruments of Different Models ......... D-29   Connecting a Series 1400a and a Partisol 2000i-D ..... D-30   Connecting a Model 2000-FRM and a Partisol 2000i/2000i-D Sampler ..................D-36  ...
  • Page 23 Figure 3–2. 47 mm Filter Placed in Bottom Section of Cassette. Top Part of Cassette Shown on Right..................3-4   Figure 3–3. Filter Cassette Carrier in Transport Container for Partisol 2000i. .. 3-4   Figure 3–4. Filter Cassette Carrier 2000i-D ............3-5  ...
  • Page 24 Figure 9–1. Front Panel Display ................9-2   Figure 9–2. Front Panel Pushbuttons ..............9-2   Figure 9–3. Partisol 2000i and 2000i-D Flowchart of Menu-Driven Firmware . 9-5   Figure 10–1. PM Inlet Showing Traditional Inlet (EPA Inlet is Similar) ..10-4  ...
  • Page 25 Mechanism ....................... 10-16   Figure 10–11. Lifting the Rollers through the Guide Slots ......10-16   Figure 10–12. Upper and Lower V-Seals on Partisol 2000i ......10-17   Figure 10–13. Upper and Lower V-Seals on Partisol 2000i-D ....... 10-17  ...
  • Page 26 Figures xxii Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 27 Descriptions ...................... 11-17   Table 11–6. I/O Expansion Board (Optional) Connector Pin Descriptions ..11-21   Table 12–1. Partisol 2000i and 2000i-D Replacement Parts ......12-3   Table 12–2. Partisol 2000i and 2000i-D Internal Cables ......... 12-7   Table 12–3. External Device Connection Components ........12-8  ...
  • Page 28 Table C–3. Write Coils for 2000i and 2000i-D ..........C-11   Table D–1. Communication Parameters ............D-31   Table E–1. Commands List .................. E-3   Table E–2. Program Register Codes ..............E-13   xxiv Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 29: Chapter 1 Introduction

    47 mm diameter sample filter(s) automatically at a user-defined time or other condition. The Partisol 2000i and 2000i-D Samplers are designed to be installed at outdoor sampling locations without additional protection from the elements.
  • Page 30 U.S. EPA. As with other manual sampling devices, the filters used in the Partisol 2000i and 2000i-D Samplers must be conditioned and weighed before exposure, and then conditioned and weighed again after use to determine the mass of the particulate matter collected during the sample exposure time.
  • Page 31: Inlet System

    Overview Figure 1–1. Partisol 2000i and 2000i-D Air Sampler Inlet System The Partisol 2000i can be configured to operate with a variety of inlet systems depending on the type of sampling being performed. For most configurations, a standard PM inlet is used, operating at a flow rate of 16.7...
  • Page 32: Internal Data Storage And Data I/O

    Advanced This section lists some of the advanced features of the Partisol 2000i and 2000i-D: Features...
  • Page 33 Automatic leak check. ● Low maintenance requirements due to durable components and a long ● life vacuum pump. Data can be retrieved using Ethernet, serial, or USB. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 34: Partisol 2000I-D Features

    Inputs received from external sources in analog voltage form are ● converted to engineering units through user-defined formulas. User input data storage rates and variables are user selectable. ● Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 35: Partisol 2000I Configuration

    ● life vacuum pump. Data can be retrieved using Ethernet, serial, or USB. ● Partisol 2000i The Partisol 2000i flow schematic provides an overview of the hardware's flow and electronic connections (Figure 1–2). The schematic shows a PM Configuration inlet that is followed by a downtube and PM cut device.
  • Page 36: Figure 1-2. Partisol 2000I Flow Schematic

    The Partisol Sampler maintains a constant volumetric flow rate at the set point entered by the user (usually 16.7 L/min), and reports sampled volumes Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 37 The mass flow sensor in the Partisol 2000i is calibrated at a temperature of 0 °C and pressure of 1 Atmosphere (1013.2 millibars or 760 mmHg). The instrument uses the measured ambient temperature and pressure to sample at the correct volumetric flow rate.
  • Page 38: Partisol 2000I-D Configuration

    U.S. EPA sampling Teflon filters are sinstalled coarse in both sample positions in the Partisol 2000i-D. For other types of sampling a range of other materials can be used in the 47 mm format, incuding TX40 (Teflon-coated glass fiber) and quartz fiber filters. Other materials are available for special applications.
  • Page 39: Figure 1-3. Partisol 2000I-D Flow Schematic

    10 microns while allowing the smaller particles to move onward toward the virtual impactor or dichotomous splitter. The dichotomous sampler's virtual impactor then separates the coarse and fine Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 1-11...
  • Page 40 The symbols in equations (1), (2), and (3) represent: = mass concentration [ g/m³] of fine particle fraction = mass concentration [ g/m³] of coarse particle fraction = mass concentration [ g/m³] of PM-10 1-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 41 760 mmHg pressure and 25 °C temperature. Flow volumes referenced internally by the Sampler to 0 °C are converted to standard conditions using the following computation: Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 1-13...
  • Page 42: Specifications

    Introduction Specifications Specifications Table 1–1 lists the Partisol 2000i and 2000i-D performance specifications. Table 1–1. Specifications Safety/Electrical Meets the following safety and electrical designations: Designations CE: EN550011 Group 1, Class B (Emissions); EN55082-1 (Immunity); EN611010-1 (Safety) ETL: UL- and CSA-equivalent approval...
  • Page 43: Chapter 2 Installation

    Partisol 2000i Sampler with WINS PM , VSCC impactor, or pass through adapter tube: inlet Filter transport container with 2 cassettes and carriers 3 rainhoods and associated hardware Sample tube Partisol stand Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 44: Partisol 2000I-D

    Null modem adapter Operating manual Quick start guide iPort package RPComm package Partisol 2000i-D Partisol 2000i-D Dichotomous Sampler with virtual impactor: Inlet Anti-static filter transport container 2 dual filter carriers 4 filter cassettes 3 rainhoods and associated hardware Sample tube...
  • Page 45: Setting Up The Sampler

    Standard Hardware Configurations Setting up the Use the following procedure to set up the Sampler. The following instructions apply to both the Partisol 2000i and Partisol 2000i-D Sampler Ambient Samplers. 1. When moving the instrument, be sure to lift the unit by using the two handles.
  • Page 46: Figure 2-2. Sampler With A Large Rainhood Installed On The Right Side Panel Of The Unit And A Small Rainhood On Left

    Insert the end of the sample tube, with the 5 cm (2 inches) machined section, into the bulkhead of the Sampler. Ensure that the tube is pushed into the enclosure through the final O-ring until it stops. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 47: Figure 2-3. Sample Tube With The Pm Inlet Installed

    Locate the two screws on the left side of the enclosure. b. Remove the two screws. Be sure to retain the washers. This will expose two holes. c. Locate the ambient temperature probe assembly in the compilation package. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 48: Figure 2-5. Ambient Temperature Probe Plugged Into The "Ambient Temperature" Connector

    Sampler labeled “Ambient Temperature” (Figure 2–5). Figure 2–4. Ambient Temperature Probe Assembly Mounted on the Enclosure Figure 2–5. Ambient Temperature Probe Plugged into the “Ambient Temperature” Connector Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 49: Partisol Stand

    9. Install 47 mm filter(s) into the Sampler in the manner described in “Filter Handling and Exchange” on page 3-1. The Partisol 2000i Sampler uses a single filter and the Partisol 2000i-D uses two filters to collect the sample. Prior to performing a sample, perform a leak check as described “External Leak...
  • Page 50: Figure 2-6. Sampler Mounted On Optional Stand

    Put this hardware together in accordance with Figure 2–7 and the “Sampler Stand Parts List” that follows. ▲ Figure 2–7. Assembly Information for the Sampler Stand Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 51 4. Fasten the right and left rails to the legs and top front and back rails with F, G and H hardware (Figure 2–7). 5. Tighten all hardware. 6. Place the Sampler onto the stand and secure using E, G, and H hardware (Figure 2–7). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 52: Hardware Considerations

    10 °C, the pump will continue running until the ambient temperature exceeds 10 °C. After instrument start up when the device is not sampling, the sample pump initiates if the ambient 2-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 53: Electronics Compartment Heater Operation

    Screen Contrast necessary to change the screen contrast if the instrument is operated at extreme temperatures or lighting conditions. Note The optimal contrast will change with changes in temperature. ▲ Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 2-11...
  • Page 54 In the Main Menu, choose Service > Screen Contrast. ● SCREEN CONTRAST: CURRENTLY: 50 % SET TO: 60 % ?  CHANGE VALUE SAVE VALUE CTRST SSET INSTS STAT 2-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 55 “Computation of Mass Concentration” on page 3-11 ● Filter Handling and The Partisol 2000i can be configured to sample a wide range of ambient PM sizes, including TSP, PM , PM , and PM . It is important that the...
  • Page 56: Chapter 3 Filter Handling And Exchange

    Pallflex TX40 (Teflon coated glass fiber) filters, quartz fiber filters and Teflon material for PM sampling. The Partisol 2000i-D is specifically designed for sampling PM and as such the use of 47-mm diameter, 2 m pore-size Teflon coarse filters must be used for U.S.
  • Page 57: Pre-Sampling Filter Weighing

    Close the filter cassette by snapping its top part onto its bottom section. Ensure that the top and bottom pieces of the cassette are pushed completely together. Figure 3–1. Positioning a 47 mm Filter on a Balance Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 58: Figure 3-2. 47 Mm Filter Placed In Bottom Section Of Cassette. Top Part Of Cassette Shown On Right

    Figure 3–3. Filter Cassette Carrier in Transport Container for Partisol 2000i. 7. If the Sampler is a Partisol 2000i-D, the filter cassette carrier will have filter storage locations for two filter cassettes. Ensure that the proper loaded filter cassette is installed in the proper position of the filter cassette carrier.
  • Page 59: Filter Exchange

    Quality Assurance Handbook, Section 8. As specified in the Quality Assurance Handbook, the transport containers must be conductive or static free. The transport container used for the Partisol 2000i is a metal container, while the container provided for the Partisol 2000i-D is made of anti-static plastic.
  • Page 60: Figure 3-5. Gripping Handle Of Filter Exchange

    (Figure 3–5 Figure 3–6) Note: These photos illustrate the procedure on the Partisol 2000i Sampler, but the procedure is identical for the 2000i-D. Figure 3–5. Gripping handle of Filter Exchange Figure 3–6. Pulling Back on Handle of Filter Exchange 4.
  • Page 61: Figure 3-7. Inserting/Removing Filter Cassette In Filter Cassette Carrier In The Partisol 2000I

    Figure 3–7. Inserting/Removing Filter Cassette in Filter Cassette Carrier in the Partisol 2000i Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 62: Figure 3-8. Inserting/Removing Filter Cassette In Filter Cassette Carrier In The Partisol 2000I-D

    Data records. Move the cursor to the type of data to view, Filter, User, or Interval, and press . Scroll through the records on screen to view the desired data record. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 63: Post Collection Equilibration

    To check for leaks in the filter cassette, check the filter for pronounced radial streaks that extend beyond the exposed area of the filter. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 64: Post Collection Weighing

    6. Document the relative humidity, temperature, and date and time of the post-collection weighing. 7. The “zero” reading of the microbalance should be verified between each filter weighing. 3-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 65: Computation Of Mass Concentration

    Sampler. For U.S. EPA PM , this must be volumetric m³. Other sampling standards require that the standard volume be used. The Partisol 2000i and Partisol 2000i-D Samplers report volumes in both the actual or standard volume. For 24-hour PM concentration averages to be valid without adjustment for U.S.
  • Page 66 Filter Handling and Exchange Computation of Mass Concentration also indicates to the user which, if any, status conditions were encountered during sampling. 3-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 67: Figure 4-1. Pm

    Components” on page ● “Partisol 2000i PM10 Components” on page ● “Converting a Partisol 2000i from a PM2.5 Sampler to a PM10 ● Sampler” on page “Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 ● Sampler” on page...
  • Page 68: Chapter 4 Partisol 2000I Inlet Conversion

    Partisol 2000i Inlet Conversion Partisol 2000i PM2.5 WINS Components Figure 4–2. Partisol 2000i Sampler with VSCC Cyclone Installed Partisol 2000i The following components comprise the PM WINS inlet configuration as shown in Figure 4–3 Figure 4–4. WINS Components Inlet WINS Impactor...
  • Page 69: Figure 4-4. Sampler With Wins Impactor Installed

    The following components comprise the PM inlet configuration as shown Figure 4–5 Figure 4–6. Components Straight Tube Inlet Sample Tube Straight Tube Adapter Figure 4–5. Straight Tube, Straight Tube Adapter, Sample Tube, and Traditional Inlet Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 70: Converting A Partisol 2000I From A Pm

    Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM2.5 Sampler to a PM10 Sampler Figure 4–6. Sampler with Straight Tube Installed Converting a Partisol Use the following procedure to convert a Partisol 2000i configured as a Sampler to a PM Sampler.
  • Page 71: Figure 4-8. Adapter For The Vscc, Wins Impactor And Straight Tube

    Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM2.5 Sampler to a PM10 Sampler 4. Allow the filter platform to settle downward, and remove the VSCC or WINS impactor and adapter by pulling them straight downward. 5. Remove the adapter (Figure 4–8) from the VSCC (or WINS impactor)
  • Page 72: Figure 4-10. Filter Exchange Rollers Shown On Top Of Lower Guide

    Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM2.5 Sampler to a PM10 Sampler Figure 4–10. Filter Exchange Rollers Shown on Top of Lower Guide 8. Install a filter carrier (with filter cassette) into the filter exchange mechanism and close the mechanism by pushing the handle toward the...
  • Page 73 Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 Sampler Figure 4–12. PM Inlet on the Sample Tube The Partisol 2000i Sampler is now configured for PM sampling. There is no difference in Sampler programming or operation with the Sampler...
  • Page 74 Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 Sampler 1. Open the Partisol 2000i enclosure door to access the filter exchange mechanism. 2. Pull the filter exchange mechanism toward you to open the filter exchange mechanism.
  • Page 75: Figure 4-13. Partisol 2000I Pm

    Partisol 2000i Inlet Conversion Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 Sampler Figure 4–13. Partisol 2000i PM with VSCC Note Note Make sure that the filter cassettes, tranport containers, and data sheets are clearly labeled as PM...
  • Page 77: Operational Modes

    Wait mode offers the user the choice of entering the Audit or Stop operating modes. Editing of sample information is not allowed in this mode. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 78: Sampling Mode

    Except in the case of Advanced sampling with conditions, the unit will draws ambient air in through its sampling system at the volumetric flow rates specified by the user. For the Partisol 2000i the default flow rate is 16.67 L/min and for the Partisol 2000i-D the default flow is 1.67 L/min for the coarse flow channel and 15 L/min for the fine...
  • Page 79: Audit Mode

    Main Menu. In the Main Menu, scroll to Audit and Calibration, and press . In the Audit and Calibration menu, scroll to Audit mode and press . At the Audit Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 80: Operation After A Power Failure

    If the hardware is not in the Sampling Operating Mode at the time of ● the power failure, the unit resumes operation in its pre-outage operating mode upon restoration of power. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 81: Sampler Settings

    ), low flow (1.67 L/min) channel located in the front sampling coarse compartment. In the software screens on the Partisol 2000i-D, the high flow (15.0 L/min), PM channel, is shown as “Flow” and the filter in the high flow sampling position is “Filter,” and the low flow (1.67 L/min), PMc channel, is “Flow 2”...
  • Page 82: Default Sample Setup

    Start Time 1 - The Start Time 1 parameter defines the time and date that the Sampler will begin sampling for the desired sample filter. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 83: Advanced Filter Method

    PRES Current ambient pressure (mmHg) %RH Current relative humidity (%) WNDSPD Current wind speed (km/h) WNDDIR Current wind direction (deg) AI1Current A/I 1 (engineering units) AI2Current A/I 2 (engineering units) Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 84: Default Filter Type

    The default flow rate parameter determines the sample flow rate (volumetric L/min). For the Partisol 2000i Sampler, the flow rate is typically 16.7 L/min. The Partisol 2000i-D Sampler uses two flow controllers to maintain the flow through the fine and coarse channels of the sampler. For the Sampler...
  • Page 85: Default Sample Times

    The default value of Average Pressure is 999, indicating that the unit should use the reading from the Sampler’s ambient pressure sensor to maintain a constant volumetric flow rate. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 86: Standard Pressure

    The Current date and time parameter are the current local date and time (or other standard time selected by the user) expressed by default as hh:mm and the current local date expressed by default as yyyy/mmm/dd. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 87: Sampler Operation

    1. For these audit procedures, prepare a set of audit cassettes with filters and solid leak check disks for use during the audit. Two cassettes are required for a complete audit of the Partisol 2000i Sampler. (Two pairs of cassettes, one each open and a blocked leak check disk, are required for the Partisol 2000i-D.)
  • Page 88: Verifying The Ambient Air Temperature

    2. Verify that the value of Amb Temp displayed in the Audit screen is within ±2 °C of the measured temperature. If the value is not within the limits, perform a calibration when the Sampler is not in Sample mode. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 89: Verifying The Filter Temperature

    Partisol 2000i and Figure 7–2 shows the Partisol 2000i-D. Figure 7–1. Filter Temperature Probe Location in Partisol 2000i Figure 7–2. Filter Temperature Probe Location in Partisol 2000i-D Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 90: Verifying The Ambient Pressure

    ▲ 1. Install the leak test filter cassettes. The Partisol 2000i requires a single filter cassette and the Partisol 2000i-D requires two filter cassettes. Carefully insert the filter cassette carrier into the lower part of the filter exchange mechanism.
  • Page 91: Figure 7-3. Filter Cassette And Carrier Installed Into Filter Exchange Mechanism In The Partisol 2000I

    Sampler and install the supplied flow audit adapter on the end of the sample tube (Figure 7–5). Turn the valve on the flow audit adapter to its closed position. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 92: Figure 7-5. Flow Audit Adapter Installed

    7. If leak check fails a second time, the leak is external to the filter. Isolate the leak and repair. 8. After a successful leak check, slowly open the valve on the flow audit adapter. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 93: Flow Verification

    2-3 minutes (minimum) thereafter. Flow Audit/Verification – Use the following procedure to perform a flow audit on the Partisol 2000i: Partisol 2000i 1. Remove the flow audit adapter used during the leak check and install...
  • Page 94: Figure 7-6. Streamline Pro Installed On External Sample Tube

    Use the following procedure to perform a flow audit on the 2000i-D: Partisol 2000i-D The audit procedure is similar to that of the Partisol 2000i single sample version of the Sampler. The main steps of the audit are the same, but must...
  • Page 95: Figure 7-7. Streamline Pro Installed On External Sample Tube

    The default sample flow should be 1.67 L/min, unless it was changed in the default settings of the Sampler. Record the flow rate from the Streamline Pro and the Sampler. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 96: Internal Leak Check

    If the system fails the leak check a second time, the leak is before the test cassette. Isolate the the leak and repair. 7-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 97: Programming A Sample Run

    A filter cassette carrier with a 47 mm filter(s) must be installed in the hardware before starting the sampling run. The filter cassette carrier used with the Partisol 2000i holds a single filter cassette while the filter cassette carrier used in the Partisol 2000i-D holds two filter cassettes.
  • Page 98 To Apply Default Times, scroll to Apply Default Times in the Sample Setup menu, press to display the Apply Default Times screen, and press to set the default times. 7-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 99 Filter2 Type screen. Use the pushbuttons as indicated on the screen to change the Filter2 Type, and press to save the setting. Press to return to the Sample Setup menu. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 7-13...
  • Page 100: Post-Sampling Verification And Data Retrieval

    6. Insert a new filter cassette assembly with a 47 mm filter into the filter exchange mechanism. Push the handle to raise the platform, and enclose the filter cassette in the sampling position and program the Sampler for the next sampling run. 7-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 101: Viewing And Downloading Stored Data

    Flow_min, Flow_ave, Flow_max - These are the minimum, average, and maximum volumetric flow rates (L/min) during the sampling period for the flow in a Partisol 2000i and the main flow in a Partisol 2000i-D. Flow_COV - The coefficient of variation is equal to the standard deviation of the five-minute flow rate averages divided by the average flow rate (16.7...
  • Page 102 (1.67 L/min by default for Flow 2 in a Partisol 2000i-D). This value is then multiplied by 100 to yield the %COV value shown on the run screen and Filter Data records.
  • Page 103 PM sample. For the Partisol 2000i, this will equal the Volume (std) and for a Partisol 2000i-D Sampler, this will equal the sum of the sample Volume std) and sample Volume2 (std). Power_fail_time(s) - At the end of the filter data records, up to ten time fields (hh.mm) that indicate the starting time of all power outages longer...
  • Page 104: Usb Data Download

    Post-Sampling Verification and Data Retrieval Flow - Average sample flow over the recording period. This is the sample flow in a Partisol 2000i and the main or PM flow in a Partisol 2000i- Flow2 - Average sample flow2 over the recording period. This is the coarse flow in a Partisol 2000i-D p-amb - Average ambient pressure over the recording period.
  • Page 105: Data Download Via Rs-232

    Filter Record 1 is the oldest record. Filter Record 2 is the most recent record of the 32 maximum filters. Press to display the different filter types. Press to display the individual records. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 7-19...
  • Page 106: Display Sampler Data

    The Status screen displays current staus conditions, but does not display status conditions from previous completed samples. Refer to Chapter 9 “Firmware Screen Descriptions” for more information on the status codes. 7-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 107: Chapter 8 Calibration

    Two cassettes are Setup required for a complete audit of the Partisol 2000i Sampler. (Two pairs of cassettes, one each open and a blocked leak test cassette, are required for the Partisol 2000i-D.)
  • Page 108: Ambient Air Temperature Calibration

    The Sampler automatically adjusts the corresponding offset based on this input. 3. If removed, reinstall the ambient temperature probe in the radiation shield, and tighten the two screws on either side of the probe. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 109: Ambient Pressure Calibration

    3. Enter the measured ambient pressure and press . The Sampler automatically adjusts the corresponding offset based upon this input. 4. If the Sampler is a Partisol 2000i-D, repeat these steps to calibrate the filter2 pressure sensor. Thermo Fisher Scientific...
  • Page 110: Filter Compartment Temperature Calibration

    3. At the Calibrate Filter Temperature screen, enter the reference temperature reading in °C [°C = 5/9 x (°F - 32)] and press save the value. The Sampler automatically adjusts the corresponding offset based upon this input. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 111: Flow Calibration

    Enter the m and b constants for the FTS sensor(s) into the Sampler. If the Sampler is a Partisol 2000i, enter a single set of constants. If the Sampler is a 2000i-D, enter constants for two flow sensors. To use the FTS as the calibration device, use the following procedure.
  • Page 112: Flow Calibration - Partisol 2000I

    Sampler. 3. Install the audit filter cassettes prepared as described earlier in this section. The Partisol 2000i requires a single filter cassette. Carefully insert the filter cassette carrier into the lower part of the filter exchange mechanism. The carrier contains a slot and hole that fit into the appropriate locations of the lower part of the filter exchange mechanism.
  • Page 113 Thermo recommends that these setpoints be used as they are configured for 10% above and below the usual setpoint. For the Partisol 2000i Sampler, the following values are recommended for a three-point calibration of the 16.7 L/min flow rate setpoint: Setpoint 1 (Min Flow): 15.0 L/min (10% below the usual 16.7...
  • Page 114: Flow Calibration - Partisol 2000I-D

    1. Make sure the Dual Flow is enabled from the calibration setup flow 2 screen. a. Go to main menu, scroll to Instrument Setup, press enter b. Scroll to Flow Calibration Setup, press enter c. Scroll to Flow 2, press enter Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 115 Use the menu key to escape back to the flow calibration menu and use the flow to verify the 20 LPM flow channel calibration settings. Operational for Flow is 15.00 LPM Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 116: Using The Flow Calibration Cassette

    Using the Flow If you do not have the Flow Calibration Cassette, skip to “Using the Solid Leak Check Disc:”. Calibration Cassette: Figure 8–1. Flow Calibrationi Cassette 8-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 117: Figure 8-2. Flow Calibration Cassette In The Front Position

    7. Place the dual filter holder in the filter exchange mechanism and close the filter exchange mechanism making sure the seals are not compromised. Figure 8–3. Dual Filter Holder in the Filter Exchange Figure 8–4. Filter Exchange Closed Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 8-11...
  • Page 118: Figure 8-5. Flow Calibration Cassette In The Back Position

    Figure 8–5. Flow Calibration Cassette in the Back Position 14. Place the dual filter holder in the filter exchange mechanism and close the filter exchange mechanism making sure the seals are not compromised. 8-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 119: Figure 8-6. Dual Filter Holder In The Filter Exchange

    17. At the end of the calibration the unit will display the calculated intercept or offset and span values necessary to linearize the flow sensor to the operational range. 18. Calibration procedure complete perform the audit per the operating manual procedure. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 8-13...
  • Page 120: Using The Solid Leak Check Disc

    Pull the bottom part of the holder out of the track allowing you room to access the flow connections on the left side of the filter holder bottom plate. 8-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 121: Figure 8-10. Pull Forward On Handle Of Filter Exchange

    Make sure you do not pinch the coarse flow tubing and that it is still open to the compartment air. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 8-15...
  • Page 122: Figure 8-12. Filter Exchange Closed

    Fine (Back) position of the holder. Move the regular cassette with filter media to the Coarse (Front) position of the holder. Figure 8–13. Solid Leak Check Cassette in the Back Position 8-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 123: Figure 8-14. Pull Forward On Handle Of Filter Exchange

    Make sure you do not pinch the coarse flow tubing and that it is still open to the compartment air. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 8-17...
  • Page 124: Figure 8-16. Filter Exchange Closed

    41. Reconnect the Fine flow tube to the front elbow fitting. Make sure the fitting is fully inerted into the elbow; you can feel it hit the stop. 42. Calibration procedure is complete. Perform flow audit per the operating manual instruction. 8-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 125: Chapter 9 Firmware Screen Descriptions

    (Figure 9–1). Some menus contain more items than can be displayed at one time. For these menus, use to move the cursor up and down to each item. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 126: Pushbuttons

    If the liquid crystal contacts your skin or clothes, wash it off immediately using soap and water. ▲ Pushbuttons The Pushbuttons allow the user to traverse the various screens/menus (Figure 9–2). Figure 9–2. Front Panel Pushbuttons Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 127: Soft Keys

    “edit soft key prompt” screen will not come up upon entering right-arrow-soft key combinations. All items under the Service menu (including the menu itself) cannot be assigned soft keys. ▲ Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 128: Alphanumeric Entry Screen

    You can customize the appearance and contents of the Run screen. See “Run Screens” on page 9-71. From the Run screen, the Main Menu can be displayed by pressing Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 129: Figure 9-3. Partisol 2000I And 2000I-D Flowchart Of Menu-Driven Firmware

    This chapter describes each submenu and screen in detail. Refer to the appropriate sections for more information. Figure 9–3. Partisol 2000i and 2000i-D Flowchart of Menu-Driven Firmware Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 130: Power-Up Screen

    The word “STOP” on the left of the status bar indicates the Sampler is in “STOP” mode. Other modes appear in the same area of the display. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 131: Custom Run Screens

    ▲ to move the cursor up and down. ● Press to make a selection. ● Press to return to the Main Menu or to return to the Run ● screen. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 132: Sample Setup Menu

    Apply EPA Times The Apply EPA Times screen is used to automatically apply midnight-to- midnight, 24-hour continuous sampling EPA times. In the Main Menu, choose Sample Setup > Apply EPA Times. ● Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 133: Apply Default Times

    The Start parameter defines the time and date that the Sampler will begin sampling for the associated filter. In the Main Menu, choose Sample Setup > Sample > Start. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 134: Filter Blank

    Yes, the filter will be a blank filter. In the Main Menu, choose Sample Setup > Sample > Filter Blank. ● FILTER IS A BLANK: CURRENTLY: SET TO: NO ? TOGGLE VALUE CTRST SSET INSTS STAT 9-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 135: Filter Id

    Filter Type is used to change the filter type, such as “P” for an EPA supplied filter. The user can select another alphabetic character to identify different filter types. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-11...
  • Page 136: Filter Type

    The Sampler has a capacity of greater than 17 days of input data at the default settings (one minute logging with 18 data items). In the Main Menu, choose View Records. ● 9-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 137: Filter

    DATE TIME Set START 2010-Oct-04 16:00 Set STOP 2010-Oct-05 16:00 Actual START 2010-Oct-04 16:54 Actual STOP 2010-Oct-04 15:05 SAMPLING: Valid = 00:05 Total = 00:05 CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-13...
  • Page 138: Format Of Filter Data Records

    Cassette ID (32 character string) Set Start Date (yyyy/mm/dd) Set Start Time ( hh:mm) Actual Start Date (yyyy/mm/dd) Actual Start Time (hh:mm) Set Stop Date (yyyy/mm/dd) Set Stop Time (hh:mm) Actual Stop Date 9-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 139 Sampled Volume 2 (m Filter Temperature Minimum 2 (°C)* Filter Temperature Average 2 (°C)* Filter Temperature Maximum 2 (°C)* Filter Temperature Difference 2* Date 2 of Max Temp Diff (yyyy/mm/dd)* Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-15...
  • Page 140: Interval/Srec

    (at one minute logging of 18 data items) before it overwrites the oldest records. Default setup: Flow Flow2* Ambient Pres Filter Pres Filter2 Pres* Pump Pres Ambient Temp Filter Temp 9-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 141: Interval/Srec By Record Number

    In the Main Menu, choose View Records > Interval/Srec by Rec No. ● SET # BACK FROM CURRENT: 0000000 TOTAL INTERVALS:  MOVE CURSOR  CHANGE VALUE SAVE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-17...
  • Page 142: User/Lrec By Record Number

    Date and Time screen. In the Main Menu, choose View Records > Interval/Srec by Date. ● 9-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 143: User/Lrec By Date

    Menu trigger method, system setup parameters, flow calibration setup, datalogging, site ID, communications, USB setup, date/time, timezone, and password. In the Main Menu, choose Instrument Setup. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-19...
  • Page 144: Default Sample Setup

    The Method screen allows the user to select the type of sampling program such as Basic (default), Time, Time 2, Advanced, and Episodic. In the Main Menu, choose Instrument Setup > Default Sample Setup > ● Method. 9-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 145: Flow Rate/Flow2 Rate

    CHANGE VALUE SAVE CTRST SSET INSTS STAT Default Sample The Default Sample Times menu is used to set the default start time (start time2 2000i-D only), duration/duration2 (2000i-D only). Times Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-21...
  • Page 146: Start Time

    STAT Trigger Method The Trigger Method menu allows the user to choose trigger of either Advanced or Episodic. In the Main Menu, choose Instrument Setup > Trigger Method. ● 9-22 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 147: Setup Trigger 1-3

    The minimum and maximum trigger screens function the same way. Trigger In the Main Menu, choose Instrument Setup > Advanced Trigger > ● Setup Trigger 1-3 > Trigger Minimum. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-23...
  • Page 148: System Setup

    The default setting for Standard Temperature is 99 °C, which may need to be changed to match conventions in different parts of the world. The default setting for Standard 9-24 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 149: Average Pressure

    In the Main Menu, choose Instrument Setup > System Setup > Std ● Pressure. SET STANDARD PRESSURE: CURRENTLY: 760 mmHg SET TO: mmHg  MOVE CURSOR  CHANGE VALUE SAVE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-25...
  • Page 150: Filter Fan

    The Flow Calibration Setup menu is used to setup flow calibration. Flow 2 is used with the 2000i-D. Setup In the Main Menu, choose Instrument Setup > Flow Calibration ● Setup. 9-26 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 151: Flow Calibration Setup

    The Set Number of Points screen allows the user to adjust the number of points. In the Main Menu, choose Instrument Setup > Flow Calibration Setup ● > Flow > No of Cal Points. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-27...
  • Page 152: Fts Constant M/B

    The Sampler’s computer system includes three megabytes of flash memory, which is enough to store a full lrec containing 32 data items and a full srec containing 32 items once each minute for a week (>20,000 total records). 9-28 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 153: Select Content

    Choices include Other Measurements and Analog Inputs (if the I/O expansion board is installed). In the Main Menu, choose Instrument Setup > Datalogging Settings > ● Select Content > Field 1–32. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-29...
  • Page 154: Other Measurements

    “Commit Content” that follows. In the Main Menu, choose Instrument Setup > Datalogging Settings > ● Select Content > select field > Analog Inputs. 9-30 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 155: Other Measurements 2

    In the Main Menu, choose Instrument Setup > Datalogging Settings > ● Select Content > select field > Other Measurements 2. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-31...
  • Page 156: Commit Content

    The Reset to Default Content screen is used to reset the logging settings to default for the selected record type. IMPORTANT Be aware that performing this action will delete all of your logged data. ▲ 9-32 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 157: Configure Datalogging

    The Memory Allocation Percent screen is used to select the percentage of Percent each record type for both srecs and lrecs. The “other” type is set to 100% which is the specified value. Percentages between 0 and 100% are available Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-33...
  • Page 158: Data Treatment

    In the Main Menu, choose Instrument Setup > Datalogging > ● Configure Datalogging > Flag Status Data. SET USER FLAG STATUS DATA: CURRENTLY: SET TO:  CHANGE VALUE SAVE CTRST SSET INSTS STAT 9-34 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 159: Erase User Data Log

    CTRST SSET INSTS STAT Site ID The Site ID menu is used to enter Sampler site location information. In the Main Menu, choose Instrument Setup > Site ID. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-35...
  • Page 160: Communications

    INSTS STAT Serial Settings The Serial Settings menu is used for serial communications control and configuration.  In the Main Menu, choose Instrument Setup > Communications > Serial Settings. 9-36 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 161: Baud Rate

    Selections of NONE, EVEN, or ODD are available, with NONE being the default value.  In the Main Menu, choose Instrument Setup > Communications > Serial Settings > Parity. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-37...
  • Page 162: Rs-232/Rs-485 Selection

    MAKE SURE THAT THE CABLEXX ● SELECTION!XXXXXX  IS OFF: PRESS TO CONFIRM  TO CONTINUE  TOGGLE VALUE ● CTRST SSET INSTS STAT CTRST SSET INSTS STAT ● 9-38 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 163: Instrument Id

    The MODBUS protocol support for the Partisol 2000i and 2000i-D enables the user to read analog values or variables, read the status of the digital outputs of the instrument, and to trigger or simulate the activation of a digital input to the instrument.
  • Page 164: Streaming Data Configuration

    The Add Labels screen allows the operator to toggle between Yes and No. If set to Yes, then each data point in the streaming record will have a label attached. 9-40 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 165: Prepend Timestamp

    In the Main Menu, choose Instrument Setup > Communications > ● Streaming Data Config > Item # > Measurements. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-41...
  • Page 166 FLOW2 COV VOLUME VOLUME2 AMBIENT PRES FILTER PRES FILTER2 PRES AMB PRES AVG FLT PRES AVG FLT2 PRS AVG AMB TMP 5MIN FLT TMP 5MIN FLOW 5MIN FLOW2 5MIN 9-42 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 167: Table 9-2. Data In Streaming Records - Other Measurements

    If the I/O expansion board is not present, Analog Inputs will not be available.  In the Main Menu, choose Instrument Setup > Communications > Streaming Data Config > Item # > Analog Inputs. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-43...
  • Page 168: Table 9-3. Data In Streaming Records - Analog Inputs

    NON- MEASUREMENTS: >NONE FILTER ID FILTER2 ID CASS ID  CASS2 ID CTRST SSET INSTS STAT START DATE STOP DATE ELAPSED TIME TIME DATE DATE/TIME EXT ALARMS STATUS CODES 9-44 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 169: Tcp/Ip Settings

     In the Main Menu, choose Instrument Setup > Communications > TCP/IP Settings. TCP/IP SETTINGS: >USE DHCP IP ADDR 192.168.1.60 NETMASK 255.255.252.0 GATEWAY 192.168.1.1  HOST NAME iseries CTRST SSET INSTS STAT NTP SVR 192.168.1.200 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-45...
  • Page 170: Ip Address

    DHCP is On, the instrument will respond with “Not Settable if DHCP is On.” For more information on DHCP, see “Use DHCP” in this chapter.  In the Main Menu, choose Instrument Setup > Communications > TCP/IP Settings > Netmask. 9-46 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 171: Default Gateway

    The Network Time Protocol (NTP) Server screen is used to edit the IP address of the NTP server. An NTP server may be used to periodically Server synchronize the instrument’s real-time clock with a standard. More Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-47...
  • Page 172: Usb Setup

     In the Main Menu, choose Instrument Setup > USB Setup > File Format. FILE FORMAT: CURRENTLY SET TO: SPACE SEP ?  CHANGE VALUE SAVE CTRST SSET INSTS STAT 9-48 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 173: Date/Time

    The Date/Time screen allows the user to view and change the system date and time (24-hour format). The internal clock is powered by its own battery when instrument power is off. In the Main Menu, choose Instrument Setup > Date/Time. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-49...
  • Page 174: Timezone For Timeserver

    The password (lock) icon on the right side of the status bar indicates that the password lock is active. The items visible under the password menu are determined by the instrument’s password status. 9-50 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 175: Set Password

    Instrument, the instrument reports being in Local mode, the front panel interface is unlocked, and settings can be changed from the front panel. In the Main Menu, choose Instrument Setup > Password > Lock ● Instrument. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-51...
  • Page 176: Unlock Instrument

    The Remove Password screen is shown if the instrument is unlocked and the password is set. In the Main Menu, choose Instrument Setup > Password > Remove ● Password. 9-52 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 177: Remove Password

    Leak checks and flow verifications can also be done with the Sampler in the Audit mode. In the Main Menu, choose Audit and Calibration > Audit Mode. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-53...
  • Page 178: Audit

    The Enter Reference Value screen is used to adjust the reference value. In the Main Menu, choose Audit and Calibration > Audit > Ambient ● Temperature > Record Ref. 9-54 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 179 View History Filter Temp Record Ref View History Ambient RH Record Ref View History Filter Pres Record Ref View History Filter2 Temp Record Ref View History Filter2 Pres Record Ref Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-55...
  • Page 180: Calibration

    Ambient Pressure The Calibrate Ambient Pressure screen displays the current and latest averaged values of the ambient pressure (mmHg), as measured by a sensor located inside the Sampler enclosure. 9-56 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 181: Filter Temperature

    In the Main Menu, choose Audit and Calibration > Calibration > Filter ● Pres/Filter2 Pres. CALIBRATE FILTER PRES: CURRENTLY: 000 mmHg SET TO: mmHg  MOVE CURSOR  CHANGE VALUE SAVE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-57...
  • Page 182: Flow

    Cal Setup The Cal Setup Flow screen provides parameters for flow calibration setup. In the Main Menu, choose Audit and Calibration > Calibration > Flow ● > Cal Setup. 9-58 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 183: Setpoint 1-5

    Reset Flow Cal menu item is used to reset the flow calibration to the default value. In the Main Menu, choose Audit and Calibration > Calibration > Flow ● > Cal Setup > Reset Flow Cal. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-59...
  • Page 184: Flow Constant A/B

    The Leak Check menu provides access to the internal leak check parameters and the external leak check parameters used in performing leak checks. In the Main Menu, choose Audit and Calibration > Calibration > Leak ● Checks. 9-60 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 185: Leak Checks

    ● Checks > External. LEAK CHECK EXTERNAL: MAKE SURE THAT FILTER IS IN PLACE AND LEAK CHECK ADAPTER IS CLOSED!!!  PRESS TO BEGIN TESTING CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-61...
  • Page 186: System Status

    In the Main Menu, choose System Status > Status Codes. ● STATUS CODES: >AMBIENT (A) FILTER TEMP (T) ELECT TEMP (E) CTRST SSET INSTS STAT 9-62 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 187: Table 9-5. Status Codes

    Coefficient of Variation 1. The coefficient of variation of the sample flow #1 rate (high flow), expressed as a percentage, was greater than 2 (2%) during the exposure of a filter. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-63...
  • Page 188: Alarms

    There are no setting screens for these alarms. In the Main Menu, choose System Status > Alarms. ● ALARMS: >ALARMS DETECTED MOTHERBOARD STATUS INTERFACE STATUS FAIL I/O EXP STATUS CTRST SSET INSTS STAT 9-64 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 189: Usb Menu

    UPDATE FIRMWARE CTRST SSET INSTS STAT Export Data Logs The Export Data Logs screen is used to export the interval/srec, user/lrec, and filter/frec data to a USB memory device. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-65...
  • Page 190 In the Main Menu, choose USB > Front USB Port > Export Data ● Logs. EXPORT DATA LOGS: >FILTER 12 RECS INTERVAL/SREC 97 RECS USER/LREC 486 RECS CTRST SSET INSTS STAT EXPORT FILTER DATA LOG: PROGRESS CTRST SSET INSTS STAT 9-66 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 191 The file name format for backup configuration 2000i is: 2000i_xxxxx_yyy.tbf Where: xxxxx = USB ID from the setup procedure yyy = file index which increments every time you save Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-67...
  • Page 192: Backup Configuration

    USB memory device. An “*” is displayed on the version number. In the following screen, the “*” indicates that 020709 is the version number. In the Main Menu, choose USB > Front USB Port > Update ● Firmware. 9-68 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 193: Service Menu

    INSTS STAT OPTIONS DISPLAY PIXEL TEST Manual Motion The Manual Motion allows the user to choose manual motion menu. In the Main Menu, choose Service > Manual Motion. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-69...
  • Page 194: Discrete Actions

    Refer to the “Hardware Configuration” section on page B-20 “C-Link Protocol Commands” Appendix for more information on this command. 9-70 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 195: Run Screens

    Run screen (1-10), and to assign a parameter to each item. EDIT RUN SCREEN: >EDIT TITLE ENABLED NUMBER OF ITEMS ITEM 1 ANALOG IN 1 ITEM 2 ANALOG IN 2  CTRST SSET INSTS STAT ITEM 10 NONE Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-71...
  • Page 196: Edit Title

    In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Enabled. ENABLE/DISABLE RUN SCREEN: CURRENTLY: SET TO: ON ? TOGGLE VALUE CTRST SSET INSTS STAT 9-72 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 197: Number Of Items

    Run screen, such as flow volume or pump pressure. Refer to Item Number described previously. In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Item number > Other Measurements. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-73...
  • Page 198: Table 9-6. Data In Run Screens - Other Measurements

    Run screen, such as flow volume or pump pressure. Refer to Item Number described previously. In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Item number > Other Measurements 2. 9-74 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 199 9–8). It allows the user to select an analog input signal (none or analog inputs 1-8) that will be included in the Run screen. If the I/O expansion board is not present, Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-75...
  • Page 200: Table 9-8. Data In Run Screen - Analog Inputs

    NON-MEASUREMENTS: >NONE FILTER ID <-- FILTER2 ID CASS ID CASS2 ID  CTRST SSET INSTS STAT START DATE STOP DATE ELAPSED TIME TIME DATE DATE/TIME EXT ALARMS STATUS CODES 9-76 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 201: I/O Configuration

    MIB STATUS I/O BD STATUS LOCAL/REMOTE  CTRST SSET INSTS STAT SERVICE In the Main Menu, choose Service > External I/O > Output Relay ● Settings > select relay (1-10). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-77...
  • Page 202: Logic State

    In the Main Menu, choose Service > External I/O > Output Relay ● Settings > Select relay 1-10 > Instrument State > Alarms. Note The I/O BD STATUS alarm is only present if the I/O expansion board is installed. ▲ 9-78 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 203 I/O is desired, an alternative means of connection is required. ▲ In the Main Menu, choose Service > External I/O > Digital Input ● Settings. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-79...
  • Page 204: Digital Input Settings

    Settings > Select relay (1-16) > Instrument Action. CHOOSE ACTION: >NONE <-- STOP START TRIGGER SAMPLE  AOUTS TO ZERO CTRST SSET INSTS STAT AOUTS TO FS EXT ALARM 1 EXT ALARM 2 EXT ALARM 3 9-80 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 205: Analog Output Configuration (Select Channel)

    The Analog Output Configuration menu displays a list of the analog output configuration choices from which the user selects the parameter to Configuration (Select adjust for the selected output channel. Configuration choices include Action) Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-81...
  • Page 206: Select Range

    In the Main Menu, choose Service > External I/O > Analog Output ● Configuration > > select channel > > Set Minimum Value or Set Maximum Value. 9-82 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 207: Table 9-9. Analog Output Zero And Full Scale Values

    Temperature Temperature Change 0 °C 10 °C Max. Temperature Change Ambient RH 100% Wind Speed 0 km/hr 180 km/hr Wind Speed Average Wind Direction 0° 360° Wind Direction Average Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-83...
  • Page 208: Table 9-10. Signal Types Group Choices

    Analog IN 2 Comprt Temp Flow Total Analog IN 3 Elect Temp Flow COV Analog IN 4 Temp Dlt Flow2 COV Analog IN 5 Temp Dlt Max Volume Analog IN 6 9-84 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 209: Analog Input Configuration

    The descriptor may be from 1 to 3 characters in length, and defaults to IN1to IN8 (user input channel number). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-85...
  • Page 210: Units

    In the Main Menu, choose Service > External I/O > Analog Input ● Configuration > select channel (1-8) > Decimal Places. DECIMAL PLACES: CURRENTLY: SET TO:  INC/DEC  SAVE VALUE CTRST SSET INSTS STAT 9-86 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 211: Number Of Table Points

    10.0 U, where U is the previously entered unit of measure. In the Main Menu, choose Service > External I/O > Analog Input ● Configuration > select channel > select point > Volts. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-87...
  • Page 212: Set Serial Number

    CTRST SSET INSTS STAT Options Menu The Options screen is used to enable/disable the I/O board presence in the instrument. In the Main Menu, choose Service > Options. ● 9-88 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 213: Enable I/O Board

    In the Main Menu, choose Service mode, and press to toggle the ● Service mode On and Off. SERVICE MODE: CURRENTLY: SET TO: ON ? TOGGLE VALUE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-89...
  • Page 214: Diagnostics

    The I/O Board item is displayed only if the optional I/O expansion board is installed. In the Main Menu, choose Diagnostics > Voltages. ● 9-90 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 215: Motherboard Voltages

    I/O expansion board. This menu is displayed only if the optional I/O expansion board is installed. In the Main Menu, choose Diagnostics > Voltages > I/O Board. ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-91...
  • Page 216: Analog Input Readings

    This will dictate whether the input is brought to high (1) or to ground (0) to call for an action. In the Main Menu, choose Diagnostics > Digital Inputs. ● 9-92 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 217: Relay States

    In the Main Menu, choose Diagnostics > Test Analog Outputs. ● TEST ANALOG OUTPUTS: >ALL VOLTAGE CHANNEL 1 VOLTAGE CHANNEL 2 VOLTAGE CHANNEL 3  VOLTAGE CHANNEL 4 CTRST SSET INSTS STAT VOLTAGE CHANNEL 6 CURRENT CHANNEL 1-6 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 9-93...
  • Page 218: Set Analog Outputs

    The Contact Information screen displays the customer service information. In the Main Menu, choose Diagnostics > Contact Information. ● CONTACT INFORMATION: CALL CENTER: 508-520-0430 WWW.THERMO.COM CTRST SSET INSTS STAT 9-94 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 219: Preventive Maintenance

    Discard any damaged cassettes. Wipe with a clean dry cloth as required. Additional filter cassettes can be ordered from Thermo Fisher Scientific (59-005923-0001). Do not expose the filter cassettes to temperatures above 50 °C. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-1...
  • Page 220: Partisol 2000I Maintenance Items

    Rebuild Kit (59-012071) (one kit is required for each rebuild), or replaced with a new pump. Partisol 2000i In addition to routine maintenance, see the following for Partisol 2000i: Maintenance Items VSCC Cyclone - Clean the VSCC every 30 days of sampling. The cyclone must be cleaned when the unit is not sampling.
  • Page 221: Cleaning The Pm

    2. Using a Phillips-blade screwdriver, remove the four pan head screws from the top of the top plate. Lift the top plate off the four, threaded, spacer standoffs and set aside. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-3...
  • Page 222 Dry and reinstall. 4. Using a general-purpose cleaner with a paper towel, clean the top plate deflector cone and internal wall surface of the acceleration assembly. 10-4 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 223: Lower Collector Assembly Cleaning And Maintenance

    1- 1/4-inch OD sample tube. 4. Clean the lower collector assembly’s threads to ensure a tight seal when the two halves are reassembled. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-5...
  • Page 224: Pm Inlet Reassembly And Reinstallation

    ● ● ● Only the PM VSCC is approved for use in a Partisol 2000i for official U.S. EPA PM sampling. Other Cyclones may be approved for PM ) sampling by other agencies. To use the Cyclone fractionator in EPA-designated reference and equipment Samplers, you must clean the fractionator every 30 days of operation.
  • Page 225: Cyclone Inlet Cleaning

    1. Remove the cyclone from its installed position in the Sampler. The remove of the cyclone is described in the Sampler audit/verification section above. 2. Remove the Partisol 2000i Sampler adapter from the cyclone. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 226 10. Perform a system leak check. 11. If in the Audit mode, at the Main Menu, select Audit and Calibration > Audit Mode, and press to turn Off the Audit mode. 10-8 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 227: Wins Pm2.5 Impactor Maintenance

    Stop in the Audit Confirmation screen, the Sampler will not resume sampling (after your audit procedures are finished). You will need to program the next sample period and will remain in Stop mode until is pressed. ▲ Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-9...
  • Page 228: Wins Impactor Cleaning

    2. Remove the impactor assembly from the bottom section of the WINS impactor (Figure 10–5). Figure 10–5. Upper and Lower Sections of the Impactor Assembly (on left and right sides of figure) 10-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 229: Figure 10-6. Coating The 37 Mm Glass Filter With Oil

    Figure 10–6. Coating the 37 mm Glass Filter with Oil 10. Place the top of the impactor assembly onto the bottom section. 11. Place the impactor assembly into the bottom section of the WINS impactor. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-11...
  • Page 230: Wins Impactor Reinstallation

    > Audit Mode, and press to turn off the Audit mode. Virtual Impactor Use the following procedures to remove, clean, and reinstall the virtual impactor that is used in the Partisol 2000i-D Sampler. Maintenance Virtual impactor maintenance and performance verification can be Partisol 2000i-D performed while in the Stop, Wait, or Sampling modes.
  • Page 231: Virtual Impactor Cleaning Partisol 2000I-D

    2. Remove the three nuts that attach the nozzle housing to the top of the Sampler and remove the nozzle housing (Figure 10–7 Figure 10– Figure 10–7. The Virtual Impactor Installed Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-13...
  • Page 232: Figure 10-8. Drawing Of Disassembled Virtual Impactor

    10. If in the Audit mode, at the Main Menu, select Audit and Calibration > Audit and press to turn Off the Audit mode. Figure 10–8. Drawing of Disassembled Virtual Impactor 10-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 233: V-Seals Cleaning And Replacement

    The V-seal cleaning and replacement procedures for the Partisol 2000i and 2000i-D are similar. The Partisol 2000i has one pair of V-seals in the filter tray and the Partisol 2000i-D has two pairs for V-seals. In addition, the 2000i-D has a V-seal in the inlet adapter.
  • Page 234: Figure 10-10. Pulling Backward On The Handle Of The Filter Exchange Mechanism

    (Figure 10–11). Figure 10–11. Lifting the Rollers through the Guide Slots 7. Allow the filter platform to settle downward. 10-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 235: Figure 10-12. Upper And Lower V-Seals On Partisol 2000I

    Figure 10–12. Upper and Lower V-Seals on Partisol 2000i Figure 10–13. Upper and Lower V-Seals on Partisol 2000i-D 10. If the Sample is a Partisol 2000i-D, the following steps show how to access the V-seal on the inlet adapter. This V-seal is hidden by the virtual impactor, which will need to be removed prior to inspecting the v-seal.
  • Page 236: In-Line Filters Replacement

    1. Turn Off the Sampler. 2. Locate the in-line filters behind the sample pump. 3. Remove and replace the filters. 4. Turn on the Sampler and perform a system leak check. 10-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 237: Service Locations

    Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 10-19...
  • Page 239: Chapter 11 Troubleshooting

    Sampler’s current state. Table 11–1 lists the Diagnostics menu items and describes their function. Refer to the “Diagnostics” on page 9-90 for detailed information about the Diagnostics menu. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-1...
  • Page 240: System Status

    Guide information and indicates the checks that you should perform if you experience an instrument problem. For additional service assistance, see “Service Locations” on page 11-22. 11-2 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 241: Board-Level Connection Diagram

    Check ribbon cable on side of display disconnected board. Board-Level Figure 11–1 Figure 11–2 can be used along with the connector pin descriptions in Table 11–3 through Table 11–6 to troubleshoot board-level Connection Diagram faults. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-3...
  • Page 242: Figure 11-1. Board-Level Connection Diagram 2000I

    Troubleshooting Board-Level Connection Diagram Figure 11–1. Board-Level Connection Diagram 2000i 11-4 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 243: Figure 11-2. Board-Level Connection Diagram 2000I-D

    Troubleshooting Board-Level Connection Diagram Figure 11–2. Board-Level Connection Diagram 2000i-D Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-5...
  • Page 244: Connector Pin Descriptions

    RS485_N to Partisol Interface Board ETHERNET ETX1_P ETX1_N ERX1_P N.C. N.C. ERX1_N N.C. N.C. I/O EXPANSION +15V BOARD +24V +24V RS485_P RS485_N INTF 24V +24V I/O EXPANSION GNDC CONNECTOR PFLT_NC DIGITAL_IN1 DIGITAL_IN2 11-6 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 245 DIGITAL_IN15 ANALOG_OUT1 ANALOG_OUT2 GNDA ANALOG_OUT5 GNDA GNDA PFLT_COM PFLT_NO GNDA DIGITAL_IN3 DIGITAL_IN4 DIGITAL_IN6 GNDA DIGITAL_IN9 DIGITAL_IN11 DIGITAL_IN12 DIGITAL_IN14 DIGITAL_IN16 GNDA ANALOG_OUT2 ANALOG_OUT4 GNDA ANALOG_OUT6 USER RS-485/RS- 1A - N.C. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-7...
  • Page 246 7B - RS485_OUT_P / RTS 8B - RS485_IN_N / CTS 9B - N.C. USB A&B VBUS_A VBUS_B USB_PORTA_N USB_PORTB_N USB_PORTA_P USB_PORTB_P USB C&D VBUS_C VBUS_D USB_PORTC_N USB_PORTD_N USB_PORTC_P USB_PORTD_P 11-8 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 247 DISABLE DISABLE MCU RESET DISABLE DIGITAL OUTPUT +15V +24V +24V RESET SPI_MISO SPI_MOSI SPI_CS2 SPI_CLK EXTERNAL EXT_RS485_N ACCESSORY EXT_RS485_P +5V 1/2A +5V 1/2A +5V 1/2A N.C. N.C. +24V1/2A +24V1/2A Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-9...
  • Page 248 Signal Description Designator +24V1/2A +24V1/2A +24V1/2A JTAG N.C. N.C. N.C. N.C. N.C. N.C. N.C. N.C. N.C. 24V MONITOR +24V LCDOUT_CLK LCDOUT_LP LCDOUT_FLM LCDOUT4 LCDOUT0 LCDOUT5 LCDOUT1 LCDOUT6 LCDOUT2 LCDOUT7 LCDOUT3 11-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 249 KEYPAD_ROW3 KEYPAD_COL2 KEYPAD_COL1 KEYPAD_COL4 KEYPAD_COL3 +24V +24V AC IN NEUTRAL LIVE EARTH AC 24V POWER NEUTRAL SUPPLY LIVE EARTH AC INTF BOARD NEUTRAL LIVE EARTH DISPLAY LCDOUT_CLK LCDOUT_LP LCDOUT_FLM Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-11...
  • Page 250 LCDOUT6 LCDOUT7 LCDOUT8 LCDOUT9 LCDOUT10 LCDOUT11 LCDOUT0 LCDOUT1 LCDOUT2 LCDOUT3 LCDOUT4 LCDOUT5 LCDOUT_ONOFF +3.3V +3.3V CONTRAST KEYPAD KEYPAD_ROW1 KEYPAD_ROW2 KEYPAD_ROW3 KEYPAD_ROW4 KEYPAD_COL1 KEYPAD_COL2 KEYPAD_COL3 KEYPAD_COL4 DISPLAY BACKLIGHT +24V 1/4A 11-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 251: Table 11-4. Backplane Board Connector Pin Descriptions

    Reference Signal Description Designator MEASUREMENT N.C. INTERFACE BOARD 1 N.C. N.C. N.C. TYPE_CODE0 TYPE_CODE1 FC1_MICROSW LIFT1_UP LIFT1_DN PUSH1_DN PUSH1_UP SHUT_EXT SHUT_RET REV_CODE FC2_MICROSW LIFT2_UP LIFT2_DN PUSH2_DN PUSH2_UP CH_15 N.C. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-13...
  • Page 252 N.C. N.C. N.C. N.C. MEASUREMENT +24V INTERFACE BOARD 2 STATUS_LAMPS +15VA PGND FILTER_COMP_FANS AGND +5VD DGND 10VA TEMP_FILTER1_SEN TEMP_FILTER2_SEN FCOMP_TEMP_SEN FCOMP_RH_SENSOR AMBTEMP_SENSOR WNDDIR_SENSOR WNDSPD_SENSOR MFS_1 MFS_2 PROP_VALVE1_PWM_P PROP_VALVE1_PWM_N PROP_VALVE2_PWM_P 11-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 253 SPARE +24V PWM VALVE 1 PROP_VALVE1_PWM_P PROP_VALVE1_PWM_N SHUTTLE PCB DGND TYPE_CODE0 TYPE_CODE1 FC1_MICROSW LIFT1_UP +5VD DGND LIFT1_DN PUSH1_DN PUSH1_UP SHUT_RET +5VD DGND SHUT_ENT REV_CODE FC2_MICROSW LIFT2_UP +5VD DGND LIFT2_DN Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-15...
  • Page 254 MFS_1 N.C. AGND +10VA FLOW1 LEAK VALVE +24V FLOW1_LEAK-VALVE AUX 24V +24V SPARE FLOW2 LEAK VALVE +24V FLOW2_LEAK-VALVE MFS 2 MFS_2 N.C. AGND +10VA AC IN AC IN NEUTRAL 11-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 255: Table 11-5. Partisol Measurement Interface Board Connector Pin Descriptions

    +5VD Table 11–5. Partisol Measurement Interface Board Connector Pin Descriptions Connector Label Reference Signal Description Designator BACKPLANE BOARD 1 N.C. N.C. N.C. N.C. TYPE_CODE0 TYPE_CODE1 FC1_MICROSW LIFT1_UP LIFT1_DN PUSH1_DN Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-17...
  • Page 256 PUSH2_UP CH_15 N.C. PROP_VALVE1_P PGND PROP_VALVE2_P N.C. N.C. N.C. N.C. N.C. N.C. N.C. N.C. N.C. BACKPLANE BOARD 2 +24V STATUS_LAMPS +15VA PGND FILTER_COMP_FANS AGND +5VD DGND 10VA TEMP_FILTER1_SEN TEMP_FILTER2_SEN 11-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 257 PRESS_VENT_VALVE FLOW1_LEAK-VALVE FLOW2_LEAK-VALVE MAG_PRESS_VALVE SHUTTLE_VALVE LIFT_PUSH_VALVE SPARE +24V RS-485 DGND RS485_P RS485_N SPARE 1 SPARE1-1 SPARE1-2 SPARE1-3 DGND POWER SUPPLY +24V_IN PGND +24V OUT +24V_IN PGND STATUS LAMP1 +24V Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-19...
  • Page 258 HEATER AC HTR_AC_FUSED NEUTRAL EARTH MODEM AC MOD_AC_FUSED NEUTRAL EARTH 12 PIN TEST +5VA EN5V +3.3VA DEN3V +10VA +15VA EN+15V EN-15V REF2.5V DGND JTAG N.C. N.C. N.C. N.C. DGND 11-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 259: Table 11-6. I/O Expansion Board (Optional) Connector Pin Descriptions

    Analog Voltage Input 4 Analog Voltage Input 5 Analog Voltage Input 6 Ground Analog Voltage Input 7 Analog Voltage Input 8 Ground Current Output Return Ground Current Output 1 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 11-21...
  • Page 260: Service Locations

    Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 11-22 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 261: Chapter 12 Servicing

    A properly grounded antistatic wrist strap must be worn while handling any internal component. If an antistatic wrist strap is not available, be sure to touch the instrument chassis before Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 12-1...
  • Page 262: Firmware Updates

    , scroll to Service mode and press to toggle the Service mode On. 3. Insert the USB memory device with the firmware update into the USB port on the Sampler. 12-2 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 263: Parts Lists

    Thermo Fisher Scientific warranty. You may purchase consumable items from Thermo Fisher Scientific by calling our customer service department at 1-866-282-0430. ▲ Table 12–1. Partisol 2000i and 2000i-D Replacement Parts Replacement Parts Part Number Electronics Enclosure w/Processor PCB, 120VAC*...
  • Page 264 Solid Filter Leak Check/Separator Disk 36-004768 Filter Cassette Screens (10) 59-005147-0010 Filter Cassette Shells (10) 59-005923-0010 Anti-Stick Rings (25) 32-003519 Upper Cassette Seal 22-010318 Lower Cassette Seal 22-005957 Kit, Flow Calibration 2000i-D 59-012290 12-4 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 265 WINS PM-2.5 Impactor* 57-004006 PM-10 Inlet/PM-2.5 SCC Inlet Combo* 59-005944 PM-2.5 SCC* 57-005896 PM-10 Inlet/PM-2.5 VSCC Inlet Combo* 59-008978 PM-2.5 VSCC* 57-008976 PM-10 Inlet/PM-1 SCC Inlet Combo* 59-007942 PM-1 SCC* 10-007742 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 12-5...
  • Page 266 Push Connect Fitting, Elbow (1/8NPT Thomas Pump) 32-008756 Pump Assembly, Gast, 120VAC 55-012135-0001 Pump Assembly, Gast, 240VAC 55-012135-0002 Push Connect Fitting, Elbow (1/4NPT Gast Pump) 32-010432 Pump, Rebuild kit, Diaphragm, Gast 59-012071 12-6 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 267: Internal Cables

    6-Pin Cable Connector 06-002454 25-Pin Wiring Adapter 06-004521-0025 Internal Cables Table 12–2 lists the internal cables for the Partisol 2000i and 2000i-D. Table 12–2. Partisol 2000i and 2000i-D Internal Cables Part Number Description 101035-00 Measurement Interface Board to 24V Power Supply...
  • Page 268: External Device Connection Components

    Cable, RS-232 or RS-485 (optional) 102888-00 Terminal Board PCB Assembly, DB37F 102891-00 Terminal Board PCB Assembly, DB37M 103084-00 Terminal Board PCB Assembly, DB25M (optional) Included with optional I/O Expansion Board in all instruments. 12-8 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 269: Terminal Block And Cable Kits

    Table 12–4. External Cables Description Cable Length DB37M to open end Six feet DB37F to open end Six feet DB25M to open end Six feet RS-232 or RS-485 Six feet Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 12-9...
  • Page 270: Filter Log

    This filter log can be used to keep track of all important readings associated with each exposed filter. Users are encouraged to make photocopies of the form on the following page or to use a similar format. 12-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 271 Servicing Filter Log Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 12-11...
  • Page 272: Service Locations

    Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 12-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 273: Chapter 13 Electronics And I/O Components

    Connectors located on the motherboard include: External connectors include: RS-232/485 Communications ● Ethernet Communications ● USB Connector ● Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 13-1...
  • Page 274: I/O Expansion Board

    Analog input (voltage) ● Digital output (TTL levels) ● Digital input (TTL levels) ● Note The sampler has spare solenoid valve drivers and I/O support for future expansion. ▲ 13-2 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 275: Analog Voltage Outputs

    Only two points are necessary for linear inputs, however a larger number of points may be used to approximate non-linear inputs. All voltage inputs have a resolution of 12 bits over the range of 0 to 10 volts. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 13-3...
  • Page 276: Digital Relay Outputs

    13-4 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 277: Ethernet Connection

    (set using a DHCP server). Up to three simultaneous connections are allowed per protocol. Any serial port protocols may be accessed over Ethernet in addition to the serial port. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual 13-5...
  • Page 278: External Accessory Connector

    The USB port is USB 2.0, full speed, 12Mbps; and low speed, 1.5Mbps and is used for interacting with a USB memory device. The USB memory device must be formatted as FAT32. 4 gB or greater is recommended. 13-6 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 279: Appendix A Warranty

    (i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault or negligence of or by Buyer, (iv) use of the Products in a manner for which Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 280 OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 281: Appendix B C-Link Protocol Commands

    C-Link Protocol Commands This appendix provides a description of the C-Link protocol commands that can be used to remotely control the Partisol 2000i and 2000i-D Samplers using a host device such as a PC or a datalogger. C-Link protocol may be used over RS-232, RS-485, or Ethernet. C-Link functions can be accessed over Ethernet using TCP/IP port 9880.
  • Page 282: Commands

    It is important that each time instrument parameters are changed, that this command be sent. If changes are not saved, they will be lost in the event of a power failure. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 283: Commands List

    Commands List Table B–1 lists the Partisol 2000i and 2000i-D Sampler C-Link protocol commands. The interface will respond to the associated command strings. Items marked with an “*” apply to the Partisol 2000i-D only. Table B–1. C-Link Protocol Commands Command Description...
  • Page 284 Reports the value for the actual flow (filter 1) in L/min flow mass Reports the value for the mass flow (filter 1) in L/min flow std Reports the value for the std flow (filter 1) in L/min Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 285 Reports/sets memory allocation for short records B-15 me (menu) Simulates pressing menu pushbutton B-18 menutext Displays the text of the menu item where the cursor is currently B-20 positioned Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 286 Reports/sets the timezone string for the network time protocol B-29 server Simulates pressing up pushbutton B-18 wind dir Reports the wind direction in degrees (0-360) wind speed Reports the wind speed in km/h Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 287: Measurements

    Wait mode and is using the Basic sampling method. Send: flags Receive: flags 00120000 Figure B–1. Flags ambient pres This command reports the current value for the ambient pressure, in mmHg. Send: ambient pres Receive: ambient pres 761.8 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 288 This command reports the value for the filter2 pressure in mmHg. Send: filter2 pres Receive: filter2 pres 734.5 filter temp This command reports the value for the filter temperature in degrees C. Send: filter temp Receive: filter temp 18.8 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 289 2.40 wind dir This command reports the wind direction in degrees (0-360). Send: wind dir Receive: wind dir 269 wind speed This command reports the wind speed in km/h. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 290: Diagnostics

    24.1 14.9 4.9 3.2 -3.2 diag volt mib This command reports the diagnostic voltage measurements on the measurement interface board. The sequence of voltages is: Positive 24, B-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 291: Datalogging

    The following example sets the data treatment for concentrations in lrec as minimum. Send: set data treatment lrec min Receive: set data treatment lrec min ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-11...
  • Page 292 "flags" command. Send: erec Receive: erec 15:45 03-01-10 flags 130000 lrec srec lrec rec num srec rec num lrxy rec num srxy rec num B-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 293 These commands set the output format for lrecs and srecs, and erec data, according to Table B–2. The following example sets the lrec output format to ASCII with text. Send: set lrec format 1 Receive: set lrec format 1 ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-13...
  • Page 294: Table B-2. Record Output Formats

    = | 1 | 5 | 15 | 30 | 60 | These commands set the lrecs and srecs logging period to value in minutes. The following example sets the lrec logging period to 15 minutes. B-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 295 For more information on how to edit the scratch pad, see the “sp field” command. The following example copies the current list in scratch pad into the lrecs list. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-15...
  • Page 296 “set copy sp to lrec”, “set copy sp to srec”, or “set copy sp to stream” commands. Send: set sp field 1 5 Receive: set sp field 1 5 ok B-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 297: Table B-3. Stream Time Values

    Attaches time stamp to streaming data string frec This command retrieves the most recent filter record. If there is no data, the command returns “NO DATA”. Send: frec Receive: frec NO DATA Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-17...
  • Page 298 = | do | down | en | enter | he | help | le | left | me | menu | ri | right | ru | run | up | 1 | 2 | 3 | 4 | B-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 299: Keys/Display

    4-bit-per-pixel gray-scale image. Also note that BMP files are upside down relative to this data, i.e. the top display line is the last line in the BMP. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-19...
  • Page 300: Hardware Configuration

    This command sets the screen’s level of contrast, according to Table B–4 The following example sets the contrast level to 50%. Send: set contrast 10 Receive: set contrast 10 ok Table B–4. Contrast Levels Contrast Level Level B-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 301 Send: set save params Receive: set save params ok time This command reports the current time (24-hour time). The following example reports that the internal time is 2:15:30 pm. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-21...
  • Page 302: Communications Configuration

    Note This command cannot be used when DHCP is on. Refer to the DHCP command that follows for additional information. ▲ Send: set addr gw 192.168.1.1 Receive: set addr gw 192.168.1.1 ok B-22 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 303 This command sets the network time protocol server address, where address consists of four numbers ranging from 0-255 inclusive, separated by “.”. Send: set addr ntp 10.209.43.237 Receive: set addr ntp 10.209.43.237 ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-23...
  • Page 304 Until you cycle the power, the address assigned by the DHCP server will still be used and reported as the current address. ▲ B-24 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 305: Table B-5. Reply Termination Formats

    This command returns “bad cmd” if no host name has been set. Send: set host name analyzer01 Receive: set host name analyzer01 ok instr name This command reports the instrument name. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-25...
  • Page 306 “ok” as if the command has been accepted and acted upon, but it will not change the instrument lock status (this is for compatibility with systems expecting and “ok” response). B-26 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 307: Table B-6. Allow Mode Command Values

    The following example sets the instrument to power up in remote/locked mode. Send: set power up mode 1 Receive: set power up mode 1 ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-27...
  • Page 308: Table B-7. Power Up Mode Values

    Table B–8. Send: set layout ack Receive: set layout ack ok B-28 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 309: I/O Configuration

    This command responds with “feature not enabled” if the I/O expansion board is not detected. Send: set analog iout range 4 1 Receive: set analog iout range 4 1 ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-29...
  • Page 310: Table B-9. Analog Current Output Range Values

    2 3 ok Table B–10. Analog Voltage Output Range Values Output Range Range 0-1 V 0-100 mV 0-10 V 0-5 V 0 [cannot be set to this, but may report] Undefined B-30 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 311 The following example sets the digital output channel 4 to state 11. Send: set dout 4 11 open Receive: set dout 4 11 open ok Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-31...
  • Page 312: Table B-11. Default Output Assignment

    The active state for digital outputs is open or closed. The active state for digital inputs is high or low. Channels that have not been assigned a variable index are not displayed. Send: list dout Receive: list dout B-32 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 313 14 p-flt2-a 16 t-amb 17 t-flt 19 t-fcmp 20 t-ecmp 21 tdelt 23 tdltx 25 rh 26 wspd 27 wdir 28 wspda 29 wvel 30 wvela 31 wdira 32 fltid Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-33...
  • Page 314 These commands set the relay logic to normally open or closed for relay number value, where value is the relay between 1 and 16. The following example sets the relay no 1 logic to normally open. B-34 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 315: Record Layout Definition

    '\n'. Valid fields are: t - parse a time specifier (2 bytes) D - parse a date specifier (3 bytes) Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-35...
  • Page 316: Format Specifier For Front-Panel Layout

    The datum identified by the value source can be printed as a string 's', hexadecimal 'x', decimal 'd', or floating point 'f', or binary 'b' number. Typically, bitfield extractions are only done for decimal or hexadecimal numbers. B-36 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 317: Alarm Information

    B- Indicates a button which pops up an input dialog prompting the user for a new value using the designated input format. The input format is specified from the 'B' through the subsequent semicolon. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-37...
  • Page 318: Examples

    If the user enters, for example, '1.234', the constructed command will be 'set no coef 1.234'. NO:21f{Code_0 Code_1 Code_2 Code_3 Code_4 Code_5 Code_6 Code_7 Code_8 Code_9 Code_10 Code_11}Lset range no %d\n B-38 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 319 The background value is taken from the 7th element, and the precision specifier is taken from the 8th. If the asterisk were not present, it would indicate instead that 8 digits after the decimal point should be displayed. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual B-39...
  • Page 321: Appendix Cmodbus Protocol

    Sampler. This is achieved by using the following supported MODBUS commands. For details of the Partisol 2000i and 2000i-D MODBUS Protocol specification, see the following topics: “Serial Communication Parameters”...
  • Page 322: Tcp Communication Parameters

    0x00 to 0xFF (Passed back in response) A Slave address is not required in MODBUS over TCP/IP because the higher-level protocols include device addressing. The unit identifier is not used by the Sampler. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 323: Error Check

    If the returned output quantity is not a multiple of eight, the remaining bits in the final data byte will be padded Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 324 Here is an example of a request and response to read outputs 2–15: Request Field Name (Hex) Function 0x01 Starting Address Hi 0x00 Starting Address Lo 0x02 Quantity of Outputs Hi 0x00 Quantity of Outputs Lo 0x0D Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 325: (0X03/0X04) Read Holding Registers / Read Input Registers

    0x0A, or binary 0000 1010. Output 15 is in the fifth bit position from the left, and output 11 is the LSB of this byte. The four remaining high order bits are zero filled. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 326 Here is an example of a request and response to read registers 10–13: Request Field Name (Hex) Function 0x03 Starting Address Hi 0x00 Starting Address Lo 0x09 No. of Registers Hi 0x00 No. of Registers Lo 0x04 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 327: (0X05) Force (Write) Single Coil

    Note This function will not work if the Sampler is in Service Level mode. ▲ Request Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000 to maximum allowed by Sampler Output Value 2 Bytes 0x0000 or 0xFF00 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 328: Modbus Addresses Supported

    Data Unit (PDU) addresses. Verify the coil number on your MODBUS master to ensure that it matches the coil number on the Sampler. ▲ Note Coil status 1 indicates active state. ▲ Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 329: Table C-1. Read Coils For 2000I And 2000I-D

    FLOW VOLUME 40003&40004 FLOW2 VOLUME 2000i-D only 40005&40006 FLOW TOTAL 40007&40008 FLOW COVARIANCE 40009&40010 FLOW2 COVARIANCE 2000i-D only 40011&40012 VOLUME 40013&40014 VOLUME2 2000i-D only 40015&40016 AMBIENT PRESSURE 40017&40018 FILTER PRESSURE Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 330 ANALOG IN 5 (I/O Expansion Board installed) 40101&40102 ANALOG IN 6 (I/O Expansion Board installed) 40103&40104 ANALOG IN 7 (I/O Expansion Board installed) 40105&40106 ANALOG IN 8 (I/O Expansion Board installed) C-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 331: Table C-3. Write Coils For 2000I And 2000I-D

    Table C–3. Write Coils for 2000i and 2000i-D Coil Number Action Triggered STOP START TRIGGER SAMPLE AOUTS TO ZERO AOUTS TO FS EXTERNAL ALARM 1 EXTERNAL ALARM 2 EXTERNAL ALARM 3 Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual C-11...
  • Page 333: Appendix D Communications Using Iport Or Rpcomm

    Refer to the iPort Instruction Manual part number 102606-00 for detailed information. RPComm Because the use of AK Protocol has been enabled on the Partisol 2000i and 2000i-D Samplers, RPComm can be used for communications with the Samplers. Because of the updated operating platform compared with the original Partisol series Samplers, not all functionality is available when using RPComm.
  • Page 334: Installing Rpcomm

    3. Press the “Next >” button when the RPComm Welcome screen appears and follow the instructions to install the program. Note Refer to the RPComm manual for more instructions on installing and using RPComm. ▲ Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 335: Updating The List Of Program Register Codes In Rpcomm

    Partisol Sampler. Select the appropriate PRC list and save it to a file. Note Be sure to remember the folder the PRC list file is saved in. ▲ 4. Disconnect from the Thermo Fisher Scientific Website. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 336: Instrument Setup For Direct Communication Using Rpcomm

    4. On the RS232 Setup screen, ensure that the proper ser port settings are selected. These include the Baud rate, data bits, parity, stop bits and Rs-232/RS-485 selection. The default settings are 9600, 8, none, and Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 337: Executing Rpcomm

    1. Initiate the RPComm software by selecting the “Start” button on your PC’s screen, highlighting “Programs,” and then highlighting “RPComm.” Choose the RPComm icon to begin executing the RPComm program (Figure D–3). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 338: Figure D-3. Executing Rpcomm

    Figure D–3. Executing RPComm 2. When RPComm begins executing, two screens are always displayed: the Communicator Main screen and the Connect List screen (Figure D–4). Figure D–4. Connect List Screen Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 339: Creating A New Connection

    (Figure D–5). Figure D–5. Connection Type Screen 2. From the Connection Type screen, select the “2000” button and choose “Settings” to display the AK Protocol Setup screen (Figure D– Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 340: Figure D-6. Ak Protocol Setup Screen

    RS232 port settings, these values must sometimes be changed. To check these values, from the Connect List screen, select the “Setup” pull down menu and choose “RS232.” Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 341: Figure D-7. Communication Port Setup Screen

    “Open Group” from the “File” pull down menu of the Communicator Main screen. A list of saved connection groups will be displayed in the Save Connection Group screen (Figure D–8). Choose the desired group and select “OK.” Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 342: Figure D-8. Save Connection Group Screen

    If the connection is not successful, or if there is no instrument attached, then the serial number area will be blank or will display “99999”. ▲ Figure D–9. Main Screen D-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 343: Downloading Stored Data

    However, if the user wishes to download data only from a particular date, for example, then the position of the storage pointer will need to be set. There are three storage pointers used in the Partisol 2000i and 2000i-D Samplers, one for each type of data (filter, interval and input).
  • Page 344: Downloading Data

    2. From the RPComm Main screen (Figure D–9), select the Download Data tab. Refer to Figure D–12 Figure D–13 for a complete description of the control buttons displayed on the Download Data screen. D-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 345: Figure D-12. Download Data Screen

    Communications Using iPort or RPComm Using RPComm Figure D–12. Download Data Screen Figure D–13. Control Buttons on the Download Data Screen Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-13...
  • Page 346: Manipulating Downloaded Data

    Windows clipboard for use in other Clipboard applications. Use the following procedure to copy data to the Windows clipboard. 1. From the Partisol 2000i Main screen (Figure D–9), select the Filter Data, Interval Data or Input Data tab to display the desired data screen (Figure D–14,...
  • Page 347: Figure D-14. Downloaded Filter Data

    Communications Using iPort or RPComm Using RPComm Figure D–14. Downloaded Filter Data Figure D–15. Downloaded Interval Data Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-15...
  • Page 348: Figure D-16. Downloaded Input Data

    Save Dialog box will appear again prompting the user for a file name for the next type of data. This will continue until all downloaded data have been saved. D-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 349: Scheduling Data Downloads

    2. Check the box next to the day(s) of the week that you want the data to be downloaded. You may choose a single day each week or any other combination of days. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-17...
  • Page 350: Figure D-18. Example Of A Scheduled Download From The Connect List Screen

    Scheduled Download box. During a scheduled download, RPComm assigns file numbers or names to the downloaded data according to the file numbers that already exist in the D-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 351: Viewing System Registers

    2. On the right portion of the Registers screen is a list of all the system registers. Using the scroll bar, examine the list of registers and place a checkmark next the registers that you wish to view. Or, if desired, select Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-19...
  • Page 352: Prcomm Modem Communications

    Modem communications can be used to remotely download data from the unit, check instrument operation and to change system variables. All iSeries instrumentation can be connected to a standard external modem for communications. D-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 353: Standard Commercial External Modem Setup

    Also, some modems limit the length of the command string that may be sent. The command string can be broken into two or more segments and sent individually. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-21...
  • Page 354: Rpcomm Setup For Use Through A Modem

    RPComm under the Windows 98 or Windows NT operating systems. ▲ 1. Set up the Sampler for direct communications. Refer to “Instrument Setup for Direct Communication” on page D-4. D-22 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 355: Figure D-20. Connect List Screen

    Edit Selected Connection icon. The Connection Type screen will then be displayed (Figure D–21). Figure D–20. Connect List Screen Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-23...
  • Page 356: Figure D-21. Connection Type Screen

    56K Voice INT PnP.” The description will vary depending on factors such as your modem type, whether it is an internal or external modem and which COM port the modem is connected to. D-24 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 357: Figure D-23. Modem Properties Screen

    “Dialing String” box. This setting will be automatically configured by the system. 14. Select the “Dialing Setup >>” button to display the Dialing Properties screen (Figure D–24). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-25...
  • Page 358: Figure D-24. Dialing Properties Screen

    “99999.” If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. ▲ D-26 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 359: Serial Switching Device Setup

    1. Set up the unit for modem communications (“Standard Commercial External Modem Setup” on page D-21). Verify that modem communications to each instrument are successful before adding the serial switching device. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-27...
  • Page 360 8. The remaining portions of the screen should have been set while establishing the modem connection (step 1). Do not modify these settings. 9. Select “OK” to exit the AK Protocol Setup screen. This will display the Connection Type screen. D-28 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 361: Multiple Instruments Of Different Models

    (Figure D–20) and select the Connection icon on the tool bar. The connection to the instrument will be initiated and the proper command codes sent. When communication is established, the Partisol 2000i Main screen will be displayed (Figure D–25). Note If the connection is successful, the unit’s serial number will be displayed at the top of the screen.
  • Page 362: Connecting A Series 1400A And A Partisol 2000I-D

    Using RPComm Connecting a Series 1400a Use the following procedure to connect a Series 1400a monitor and a Partisol 2000i-D to a serial switching device. and a Partisol 2000i-D 1. Set up the 1400a monitor and Sampler for direct communications (“Instrument Setup for Direct...
  • Page 363: Table D-1. Communication Parameters

    11. Attach a phone line to the modem. 12. With the RPComm Connect List screen displayed (Figure D–20), select the name of one of the connections used to verify the direct Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-31...
  • Page 364 “Dialing String” box. This setting will be automatically configured by the system. 21. Select the “Dialing Setup >>” button to display the Dialing Properties screen (Figure D–24). D-32 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 365 D–20) and then select the Connection icon on the tool bar. The modem connection will now be initiated. When communication is established, the instrument’s main screen will be displayed. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-33...
  • Page 366 The Connection Type screen will then be displayed (Figure D–21). 35. From the Connection Type screen, ensure that the correct instrument type is selected and select “Settings” to display the AK Protocol Setup screen (Figure D–22). D-34 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 367 If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-35...
  • Page 368: Connecting A Model 2000-Frm And A Partisol 2000I/2000I-D Sampler

    Using RPComm Connecting a Model 2000- Use the following procedure to connect a Partisol Model 2000-FRM Sampler and a Partisol 2000i/2000i-D Sampler to a serial switching device. FRM and a Partisol 2000i/2000i-D Sampler 1. Set up the Partisol 2000-FRM Sampler and Partisol 2000i/2000i-D Sampler for direct communications (“Instrument Setup for Direct...
  • Page 369 Connection Type screen will then be displayed (Figure D–21). 13. From the Connection Type screen, ensure that the correct instrument type is selected and select “Settings” to display the AK Protocol Setup screen (Figure D–22). Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-37...
  • Page 370 If the phone number is not correct, there is an error on this screen or you may have incorrectly entered the phone number on the AK Protocol Setup screen. D-38 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 371: Figure D-27. Communication Port Setup Screen

    25. From the Communicator Main screen (which is located just behind the Connect List screen), select the “Setup” pull down menu and choose “RS232.” The Communication Port Setup screen will appear. The settings on this screen should match the Partisol 2000i/2000i-D Sampler settings as shown in Figure D–27.
  • Page 372 Connection Command box. If a code is entered incorrectly, it can be deleted by selecting the code in the Connection Command box and pressing the Delete key on your computer’s keyboard. D-40 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 373 If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual D-41...
  • Page 375: Appendix Eak Protocol

    AK Protocol This appendix provides a description of the AK protocol commands that can be used to remotely control a Partisol 2000i/2000i-D using a host device such as a PC or a datalogger. AK protocol may be used over RS-232, RS-485, or Ethernet.
  • Page 376: Basic Command Structure

    ‘SE’ indicates that at least one of the command’s parameters is out of range or another error has occurred. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 377: Commands List

    AK storage pointer resides at the oldest record in the internal data logger. The current record type affected by the SSTO command is denoted by PRC register 31. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 378: Table E-1. Commands List

    Up to 3 digits appended to the end of the response transmission according to <ETX> the entry for RS-Para 3. <ETX> <CR> <space> <LF> ASCII code 003. <ETX> <CR> <LF> Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 379 PRC is less than 3 characters long. <space> <CR> Space. <ETX> <LF> ASCII code 003. Current value of the variable referenced by the Ask Register command. NOTE: This value can be of varying length. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 380 Description No Err Error Description <ETX> ASCII code 003. <CR> Up to 3 digits appended to the end of <LF> the response transmission, according to the entry for RS-Para 3. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 381 Interval Data 2 in PRC 31 Input Data The instrument then transmits the number of storage records shown in response bytes 10 to 12 above. Each record is followed by <CR><LF>. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 382 RS-Para 3. variable referenced by Program Register Code in bytes 11 to 13 above. NOTE: The value entered may be of varying length. <ETX> ASCII code 003. Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 383 ASCII code 003. <ETX> ASCII code 003. <CR> Up to 3 digits appended to the end of the response transmission according to <LF> the entry for RS-Para 3. <ETX> <CR> <LF> Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual...
  • Page 384 LISTING OF FUNCTION CODES (xx) <LF> of the response transmission, according to the entry for RS-Para 3. <ETX> Stop Set Time <CR> Set Date Switch to "None" RS232 Mode <LF> E-10 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 385 250 to move forward by n records, and negative numbers <LF> such as -1000 to move backwards by n records. Do not right-fill. <ETX> ASCII code 003. Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-11...
  • Page 386 RS-Para 3. Up to 3 digits appended to the end <LF> of the response transmission, according to the entry for RS-Para 3. E-12 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 387: Program Register Codes

    Analog Output Current Channel 5 Code* Anytime Analog Output Current Channel 6 Code* Anytime Storage Download Type Code* 0 -2 Anytime Analog Input Reading 1 User defined User defined None Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-13...
  • Page 388 Digital Output PRC 9 Code* Anytime Digital Output PRC 10 Code* Anytime Digital Output Logic 1 Code* Anytime Digital Output Logic 2 Code* Anytime Digital Output Logic 3 Code* Anytime E-14 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 389 Average Filter Temperature 1 °C Never Average Filter Temperature 2 °C Never Current Wind Direction Never Average Wind Direction Never Current Wind Speed km/h 0 -180 Never Average Wind Speed km/h Never Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-15...
  • Page 390 Anytime Sample Filter Type A2 0 -26 Anytime Sample Filter Type A3 0 -26 Anytime Sample Filter Type A4 0 -26 Anytime Sample Filter Type A5 0 -26 Anytime E-16 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 391 Anytime Sample Filter Type B4 0 -9999999 Anytime Sample Filter Type B5 0 -9999999 Anytime Sample Filter Type B6 0 -9999999 Anytime Sample Filter Type B7 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-17...
  • Page 392 Anytime Sample Filter ID A6 0 -9999999 Anytime Sample Filter ID A7 0 -9999999 Anytime Sample Filter ID A8 0 -9999999 Anytime Sample Filter ID A9 0 -9999999 Anytime E-18 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 393 Anytime Sample Filter ID B8 0 -9999999 Anytime Sample Filter ID B9 0 -9999999 Anytime Sample Filter ID B10 0 -9999999 Anytime Sample Filter ID B11 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-19...
  • Page 394 Sample Filter Cassette ID A10 0 -9999999 Anytime Sample Filter Cassette ID A11 0 -9999999 Anytime Sample Filter Cassette ID A12 0 -9999999 Anytime Sample Filter Cassette ID A13 0 -9999999 Anytime E-20 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 395 Sample Filter Cassette ID B12 0 -9999999 Anytime Sample Filter Cassette ID B13 0 -9999999 Anytime Sample Filter Cassette ID B14 0 -9999999 Anytime Sample Filter Cassette ID B15 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-21...
  • Page 396 Sample Filter Blank A14 0 - 1 Anytime Sample Filter Blank A15 0 - 1 Anytime Sample Filter Blank A16 0 - 1 Anytime Sample Filter Blank A17 0 - 1 Anytime E-22 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 397 None Current Cassette 1 ID None Current Cassette 2 ID None Current Sample A Blank None Sample 1 Volume None Sample 2 Volume None Sample 1 Volume Standard None Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-23...
  • Page 398: Prc Values Defined By Codes

    Note A status code value showing “00000001” indicates that there are no status codes (NONE). ▲ PRC 5: Status Code (H)1 NONE (H)2 MEMORY (H)10 POWER FAILURE (H)20 FLOW (H)40 FLOW2 E-24 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 399 AMB PRES AVG FLT PRES AVG FLT2 PRES AVG PMP PRES AVG AMBIENT TEMP FILTER TEMP FILTER2 TEMP COMPRT TEMP ELECT TEMP TEMP DLT TEMP2 DLT TEMP DLT MAX Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-25...
  • Page 400 ANALOG IN 8 PRC 31: Storage Download Type FILTER DATA INTERVAL DATA INPUT DATA PRC 60 – PRC 65: Analog Voltage Output Range 1 volt 100 millivolt 10 volt 5 volt E-26 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 401 SERVICE EXTERNAL ALARM 1 EXTERNAL ALARM 2 EXTERNAL ALARM 3 PRC 82 – PRC 91: Digital Output Active State NORMALLY CLOSED NORMALLY OPEN PRC 162: Default Filter Type Space Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-27...
  • Page 402 AI8 NOW AI1 AVE AI2 AVE AI3 AVE AI4 AVE AI5 AVE AI6 AVE AI7 AVE AI8 AVE DIG IN1 DIG IN2 DIG IN3 DIG IN4 DIG IN5 DIG IN6 E-28 Partisol 2000i and Partisol 2000i-D Instruction Manual Thermo Fisher Scientific...
  • Page 403 DIG IN8 PRC 434: Sampling Action Under Serial Control Stop Sampling (Sampling Not Active) Exchange Cassette, Goes to State 2 (Sampling Active) Activate Sampling (Sampling Active) PRC 437: Auto Run Thermo Fisher Scientific Partisol 2000i and Partisol 2000i-D Instruction Manual E-29...

This manual is also suitable for:

Partisol 2000i-d

Table of Contents