In-Line Flow Meter Filter With Fittings Replacement - Thermo Scientific MIE ADR-1500 Instruction Manual

Particulate monitor

Advertisement

Maintenance and Service
In-Line Flow Meter
Filter with Fittings
Replacement
5-26
MIE ADR-1500 Instruction Manual
Use the following procedure to replace the in-line flow meter filter with
fittings
(Figure
5–11).
1. Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2. Unfasten the two captive hardware and swing cover down on the center
panel.
3. Disconnect the battery cable from J6 or J8 from the power
communication board.
4. Use needle nose pliers and release the fitting from the tubing.
5. To install the new filter, follow the previous steps in reverse order.
Note When replacing the new filter make sure filter arrow is facing
towards the right.