Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument

Source Interface Board and Heat Shield

364
TSQ 9610 Hardware Manual
To replace the source interface board and heat shield
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
139.
2. If you are only replacing the heat shield, you do not need to remove the analyzer tray.
Disconnect the filament board and wires from the source interface board and the left
screw to remove the heat shield.
Figure 404. Replacing the Heat Shield
Filament Board
Left
and Wires
Screw
Note Keep all manifold parts you remove covered on a clean work surface so they will not
accumulate dust.
3. Replace the heat shield.
4. Reconnect the filament board and wires to the source interface board.
5. If you are replacing the source interface board, remove the analyzer tray from the
instrument by following the instructions in
9610 Instrument"
on
Source Interface
Board
"Removing the Analyzer Tray from the TSQ
page
141.
"Shutting Down the TSQ 9610 System"
Heat Shield
Thermo Scientific