Interpreting Integrity Indicator Values - Agilent Technologies 8960 10 Series Reference Manual

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

PVT Troubleshooting

Interpreting Integrity Indicator values

See
"Integrity Indicator" on page
If over range (5) is returned; the PvT input power has exceeded the test set's internal sampler maximum value
during some part of the sampling or the input power has exceeded the calibrated range of the test set's power
detector for the RF Range setting. RF Range is automatically set based on the input power setting. Input
power is a combination of amplitude offset and expected power settings. See
If the signal has both over range and under range conditions only the over range (5) will be indicated.
If under range (6) is returned; the PvT Transmit Power result is more than 10 dB below the expected input
power level. Under range is also indicated if, the input power is more than 2 dB below the calibrated range of
the test set's power detector for the RF Range setting. RF Range is automatically set based on the input power
setting. Input power is a combination of amplitude offset and expected power settings. See
on page
564.
Syn Not Found (11) will be returned if the measurement Burst Synchronization is set to Midamble
synchronized and Expected Burst pattern is not set to TSC0 through TSC7, or RACH. see
page 239
96
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\ts_pvt.fm
125.
"Receiver example" on page
"Receiver example"
"CALL:BURSt" on
564.