CALL:COUNt:RACH?
Function
Queries the RACH counter. The RACH counter keeps track of the number of RACH bursts
received by the base station emulator during call setup attempts.
Query
Range: 0 to 9999
Resolution: 1
*RST Setting
0
Programming Example
OUTPUT 714;"CALL:COUNT:RACH?"
CALL:COUNt:TDMA:FRAMes
Function
Sets/queries the Max Frames Allowed for Assignment field. The Max Frames Allowed for
Assignment field specifies the maximum number of TDMA frames the mobile station is allowed
to take, from the start of the assignment or handover command, for a channel assignment. This is
only applicable to changes in TCH band, TCH ARFCN or TCH timeslot. Changes to any other
TCH parameter will not cause an error to be generated if the number of frames taken to perform
the change exceeds the setting of the Max Frames Allowed for Assignment field.
Setting
Range: 15 to 999
Resolution: 1
Query
Range: 15 to 999
Resolution: 1
*RST Setting
28
Programming Example
OUTPUT 714;"CALL:COUNT:TDMA:FRAMES 15" !Sets the number of TDMA frames allowed
!before transmitting the new TCH.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_count.fm
CALL:COUNt
245