Preset Descriptions
command is not recommended for use at this time.
Full Preset
A full preset requires the user to setup new measurements and their parameters. If new setup parameters are
not needed, use a partial preset to save time. Transmit power is set to its default value. Transmit power is not
set to OFF during a full preset.
Full preset behavior is the same as partial preset behavior with the exception of Trigger Arm and
Measurement Parameters, see the results listed below.
Press the blue SHIFT key and then the green PRESET key to perform a full preset.
Table 2. Full Preset Behavior
Function
Trigger Arm
Trigger Arm
Maskable Message Display
State
Maskable Message Display
State
Measurement Parameters
Example
OUTPUT 714;"*RST" !Command for a full preset when user in remote operation.
The *RST common command is the recommended command for a Full Preset operation. The
SYSTEM:PRESET2 command is not recommended for use at this time.
Status Preset
The STATUS:PRESET command will set the status system as defined in "SCPI 1995 Volume 2: Command
Reference" section 20.7. All of the enable registers will be set to 0, all PTR registers will be set to 1, and all
NTR registers will be set to 0.
Example
OUTPUT 714;"STATUS:PRESET" !Presets the STATus subsystem.
Related Topics
*******************************************************
"SYSTem:PRESet" on page 493
"*RST" on page 497
*******************************************************
536
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\gen_inst_states.fm
Result
Continuous (manual full
preset)
Single (remote full preset)
On (manual full preset)
Off (remote full preset)
all set to defaults