SETup:IQTuning
SETup:IQTuning:COUNt[:SNUMBer]
Function
Sets/queries the number of I/Q Tuning multi-measurements the test set makes. This command
sets the count state to ON.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST
10
Programming Example
OUTPUT 714;"SETUP:IQTUNING:COUNT:SNUMBER 25" !Sets the state to ON and the
multi-measurement count value to 25.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
SETup:IQTuning:REFerence[:FREQuency]
Function
Sets/queries the reference offset frequency to be used for the measurement. This means that if
your mobile is transmitting all 1s you should set this command to NEG67KHZ, and if your
mobile is transmitting all 0s it should be set to POS67KHZ. Alternatively you could select AUTO
which allows the test set to select the most appropriate offset.
Setting
NEG67KHZ|ZEROKHZ|POS67KHZ|AUTO
Query
NEG67KHZ|ZEROKHZ|POS67KHZ|AUTO
*RST
AUTO
Programming Example
OUTPUT 714;"SETUP:IQTUNING:REFERENCE:FREQUENCY NEG67KHZ" !Sets the I/Q
Tuning measurements reference frequency to -67 kHz.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX