Agilent Technologies 8960 10 Series Reference Manual page 427

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:PVTime:BSYNc
Function
Sets/queries the burst synchronization mode for power versus time measurements. See
Synchronization of Measurements" on page
Setting
MIDamble|AMPLitude|NONE
Query
MID | AMPL | NONE
*RST Setting
MID
Programming Example
OUTPUT 714;"SETUP:PVTIME:BSYNC MIDAMBLE" !Selects burst synchronization to on
SETup:PVTime:CONTinuous
Function
Sets/queries the trigger state for power versus time measurements. See
page 150
Range
Single trigger mode = 0|OFF
Continuous trigger mode = 1|ON
Query
0|1
*RST Setting
0|OFF
Programming Example
OUTPUT 714;"SETUP:PVTIME: CONTINUOUS OFF" !Specifies single trigger mode for
SETup:PVTime:COUNt[:SNUMber]
Function
Sets/queries the number of power versus time measurements the test set will make and turns the
multi-measurement count state on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:PVTIME:COUNT:SNUMBER 25" !Sets the state to on and the
115.
!midamble for power versus time
!measurements.
!power versus time measurements.
!multi-measurement count value to 25.
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pvtime.fm
SETup:PVTime
"Burst
"Measurement States" on
427

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5515aE5515bE1960a

Table of Contents