DISPlay
DISPlay:MODE
Function
Sets/queries the test set's display mode.
See
"Display Mode (Track/Fast)" on page
Range
FAST|TRACK
Query
FAST | TRAC
*RST setting
TRACK
Programming Example
OUTPUT 714;"DISPLAY:MODE FAST" !Sets display mode to fast.
DISPlay:MESSage:MASKable:STATe
Function
Blocks maskable messages from appearing on the test set display display screen but not from the
Message Log. Maskable messages are reported to the Message Log in either state.
Setting
Range: On | Off
Query
Range: On | Off
Factory
On
setting
Programming Example
OUTPUT 714;"DISPLAY:MESSAGE:MASKABLE:STATE OFF" !Prevents certain messages from
appearing on the display.
DISPlay:WINDow:ERRor:CLEar
Function
Clears the error message from the display screen but not from the Message Log.
Programming Example
OUTPUT 714;"DISPLAY:WINDOW:ERROR:CLEAR" !Clears an error message from the display.